Chapter 4 Command Reference
Introduction
Programming Guide
4-55
FETCh[[:SCALar]:<function>]?
This query returns the measurement taken by the :INITiate
(or :MEASure query or :READ?) commands.
When [:SCALar]:<function> is specified, the instrument will retrieve the specified
result if it matches the current measurement type or can be derived from the
current measurement type. The only functions which can be derived from a
different measurement type are:
•
frequency to/from period,
•
voltage minimum to/from voltage maximum,
•
voltage minimum to/from voltage peak-to-peak, and
•
voltage maximum to/from voltage peak-to-peak.
When [:SCALar]:<function> is omitted, the function specified/used by the last
:CONFigure, :MEASure, :READ, or FETCh will be used,
if possible. This behavior is apparent when switching between frequency and
period, or when switching among the voltage peaks functions.
Issuing this query while a measurement is in progress has the effect of holding off
further commands from being processed until the measurement completes. This
hold-off action can only be canceled by the measurement completing, Device
Clear, or power-on.
•
Result will be formatted according to :FORMat[:DATA] ASCii | REAL
setting.
•
When ASCii format is used, numeric data is transferred as ASCII bytes in
<NR3> format. The number of significant digits will range from 1 to 15,
depending on the measurement resolution.
•
If no valid results exists, Not a Number
9.91E37
is returned and error -230 is
generated.
•
If :FETCh? is executed while an automatically armed Totalize measurement
is
in progress
, Not a Number
9.91E37
is returned and error -221 is generated.
:ABORt should be used to terminate the measurement and avoid this error.
Query Response
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...