:CALCulate[1] Subsystem
Performs post-acquisition math (scale/offset) processing (on the data acquired by
a SENSe function) and data transfer of the scaled/offset result. See the :TRACe
subsystem for commands used to set the scale and offset.
Not until :CALCulate[1]:MATH:STATe is set to ON will any of the
:CALCulate[1] settings or :TRACe[:DATA] settings be used.
:CALCulate[1]:DATA?
Queries the current scaled and offset measurement result .
•
Result will be formatted according to :FORMat[:DATA] ASCii | REAL
setting.
•
When ASCii format is used, numeric data is transferred as ASCII bytes in
<NR3> format. The number of significant digits will range from 1 to 15,
depending on the measurement resolution. Only significant digits will be
returned.
•
If no valid result exists, Not a Number
9.91E37
is returned and error -230 is
generated.
•
If the current measurement is Totalize or Voltage Peaks, Not a
Number
9.91E37
is returned and error -221 is generated.
•
Query only.
•
If this command is issued when math is enabled and while a measurement is in
progress, no response will be produced until the measurement completes.
•
This command holds off subsequent commands from being processed until a
measurement completes. This holdoff action can only be canceled by the
measurement completing, a device clear, or power cycle.
•
The last calculated result remains valid until a new computation is made or a
relevant instrument state is modified.
NOTE
Query Response
Comments
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...