Chapter 4 Command Reference
Introduction
Programming Guide
4-31
Device Clear
Device Clear
The full capability of the Device Clear IEEE 488.1 interface function is
implemented in the Counter. This function allows a device to be initialized to a
cleared state. The device-dependent effect is described below.
In response to either the Device Clear message or the Selected Device Clear
message, the Counter:
•
clears the input buffer and Output Queue,
•
resets the parser, execution control, and response formatter,
•
clears any command that would prevent processing a *RST or other
commands,
•
disables the effect of a prior *OPC command, and
•
terminates the holdoff action of a *WAI, *OPC?, or data query (:MEASure
query, :READ query, :FETCh query, :CALC:DATA?, :CALC2:LIM:FAIL?)
waiting for pending operation to complete.
Also, a front-panel initiated diagnostic or calibration may be aborted (for example,
if the front-panel diagnostic or calibration is waiting for user input).
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...