Chapter 4 Command Reference
Introduction
Programming Guide
4-133
*OPC?
(Operation Complete Query)
*OPC?
This query produces a response upon the transition of the measurement cycle from
measuring to idle. This allows synchronization between a controller and the
instrument using the MAV bit in the Status Byte Register or a read of the Output
Queue. (Note that this query does not actually “read” a state, as most queries do.)
Since this query will not respond until the measurement cycle transitions from
measuring to idle, the only way to cancel the query “holdoff” is by Device Clear
(page 4-31) or power-on.
See the section titled “Using the *OPC? Command,” page 3- 47,
in Chapter 3 for an example using this command.
Single ASCII-encoded byte, 1.
The *OPC? query does not in any way affect the OPC bit in the Standard Event
Status Register.
Query Response
NOTE
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...