Chapter 4 Command Reference
Introduction
Programming Guide
4-109
:STATus:OPERation:NTRansition
<non-decimal numeric> | <NRf>
Sets or queries the negative transition filter for the Operation status reporting
structure.
The parameter and query response value, when rounded to an integer value and
expressed in base 2 (binary), represents the bit values of the negative transition
filter. The value of unused bits is zero when queried and ignored when set.
A TRUE bit (in the negative transition filter) specifies that a negative (TRUE to
FALSE) transition of the corresponding bit in the Operation Condition Status
Register generates the corresponding event in the Operation Event Status Register.
The range of the <non-decimal numeric> or <NRf> parameter is
0 to 65,535.
Numeric data transferred as ASCII bytes in <NR1> format.
•
At power-on and STAT:PRES, the negative transition filter is preset such that
each bit is a 0 (FALSE).
•
This value is unaffected by *RST and save/recall.
:STATus:OPERation:PTRansition
<non-decimal numeric> | <NRf>
Sets or queries the positive transition filter for the Operation status reporting
structure.
The parameter and query response value, when rounded to an integer value and
expressed in base 2 (binary), represents the bit values of the positive transition
filter. The value of unused bits is zero when queried and ignored when set.
A TRUE bit (in the positive transition filter) specifies that a positive (FALSE to
TRUE) transition of the corresponding bit in the Operation Condition Status
Register generates the corresponding event in the Operation Event Status Register.
The range of the <non-decimal numeric> or <NRf> parameters is
0 to 65,535.
Numeric data transferred as ASCII bytes in <NR1> format.
Range
Query Response
Comments
Range
Query Response
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...