Chapter 4 Command Reference
:CALCulate[1] Subsystem
4-10
Programming Guide
:CALCulate[1]:MATH[:EXPRession][:DEFine]?
Queries equation used for math operation.
A sequence of ASCII-encoded bytes:
(“SENS” * SCALE + OFFSET)
terminated with a new line and EOI.
•
Query only.
•
This query should be the last query in a terminated program message;
otherwise, error -440 is generated.
:CALCulate[1]:MATH[:EXPRession]:NAME SCALE_OFFSET
or
:CALCulate[1]:MATH[:EXPRession]:SELect SCALE_OFFSET
Sets or queries the name of the expression selected for math processing .
A sequence of ASCII-encoded bytes: SCALE_OFFSET
*RST: SCALE_OFFSET
:CALCulate[1]:MATH:STATe <Boolean>
Sets or queries the math enable.
This enable specifies whether or not measurement (SENSe) data will be scaled and
offset.
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
•
*RST: OFF
•
Updating the math enable causes the limit counts (:CALC2:LIM:FCO,
:CALC2:LIM:PCO) to be cleared.
Scale & Offset
Query Response
Comments
Query Response
Comments
Query Response
Comments
Related
Front-Panel
Keys
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...