Chapter 4 Command Reference
[:SENSe] Subsystem
4-80
Programming Guide
[:SENSe] Subsystem
The [:SENSe] subsystem commands are divided into several sections. Each
section or subtree deals with controls that directly affect
instrument-specific settings and not those related to the signal-oriented
characteristics.
[:SENSe]:DATA? [“[:]SENSe[1]”]
Queries the current measurement result data of the :SENSe subsystem (no scale or
offset applied).
If this query executes while a measurement is in progress, then the prior
measurement result will be returned, if the prior result has not been invalidated.
•
Result will be formatted according to :FORMat[:DATA] ASCii | REAL
setting.
•
When ASCii format is used, numeric data is transferred as ASCII bytes in
<NR3> format. The number of significant digits will range from 1 to 15
depending on the measurement resolution.
•
If no valid result exists, Not a Number
9.91E37
is returned and error -230 is
generated.
•
If this query is executed while an automatically armed Totalize measurement
is in progress
, Not a Number
9.91E37
is returned and error -221 is
generated. :ABORt should be used to terminate the measurement and avoid
this error.
Query only.
[:SENSe]:EVENt[1|2] Subtree
This subtree defines the “trigger event.”
For Frequency, this is the event which is counted; for Time Interval, this is the
start/stop edge.
These trigger events are fed to the sensor function. In separate, the EVENt1
subtree is fed the INPut1 signal and the :EVENt2 subtree is fed the INPut2 signal;
in common, the INPut1 signal is fed to both the EVENt1 and :EVENt2 subtrees.
When you are measuring Voltage Peaks, none of the :SENSe:EVENt settings are
used.
Query Response
Comments
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...