6 Troubleshooting
59
6.2.2 Low sensitivity
6.2.3 Noise and reading variations
6.2.4 Slow response speed
Cause
Remedial action
Reference
No gas is flowing.
Gas flow rate is insufficient.
Check whether tubes to the analyzer is choked
or disconnected.
―
Apply pressure to the sample gas and
reference gas inlets.
―
Check that the main valve of the span gas
cylinder is open.
―
Check the flow rate.
Check whether the filters are dirty. If they are,
replace them.
Reference gas contains impurities.
If impurities are found to be contained in the
gas after checking with the zero gas purifier
you provided, replace the reference gas.
―
Span gas is incorrect.
Use a span gas that is suitable for the
component to be measured.
―
Gas is leaking.
Replace the joints or otherwise repair the
piping.
―
Others Contact
HORIBA
or local representative.
―
Cause
Remedial action
Reference
Analysis part is not attached to the
analysis part base properly.
Reattach the analysis part.
Springs suspending the analysis part
are not attached properly.
Adjust or replace the analysis part suspension
fixtures.
Gas is leaking.
Replace the joints or otherwise repair the
piping.
―
Vibration levels of the installation
location are high.
Use the analyzer in a location with little
vibration.
―
Calibration span gas concentration
setting and cylinder gas concentration
do not match.
Set the span gas concentration correctly and
then perform the calibration again.
Others Contact
HORIBA
or local representative.
―
Cause
Remedial action
Reference
Sample gas, reference gas, or span
gas is not flowing at the specified flow
rate.
Adjust the flow rate.
Gas is leaking.
Replace the joints or otherwise repair the
piping.
―
Others Contact
HORIBA
or local representative.
―
Содержание GA-370
Страница 1: ...Trace Gas Monitor GA 370 Instruction Manual eng CODE GZ0000456936B...
Страница 91: ...8 Reference 75 8 5 Related diagrams Name No GA 370 External Dimensions GZ0000471868B Flow Sheet GZ0000465573...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...