3 Operations
19
Table 9 Various names and functions (CALIBRATION screen)
No.
Name
Description
Reference
1
[MODE] button
Switches between the MEASURE mode, CALIBRATION
mode (ZERO, SPAN of each component) and DELAY
mode. When performing calibration, switch to calibration
mode and perform calibration.
Switching to calibration mode allows you to switch to the
ZERO CALIBRATION screen and SPAN CALIBRATION
screen.
Furthermore, the color of the bar changes from blue to
pink.
2
Measurement values
Displays the current measurement values.
―
3 RANGE
Displays the measurement range or calibration range of
each component.
The measurement range is displayed when the
MEASUREMENT mode or DELAY mode is selected with
the [MODE] button, and the calibration range is displayed
when the CALIBRATION mode is selected.
4
SPAN GAS
Displays the entered span gas concentration value.
5
[BACK] button
Returns to the CALIBRATION MENU screen.
―
6
Calibration alarm cancel
button
Displayed in each calibration screen when a calibration
alarm is occurring. Pressing this button cancels the error.
After pressing this button to cancel a calibration alarm,
perform calibration again.
7
buttons
Switches the screen.
―
8 [ZERO]
button
Executes
zero calibration.
9 [SPAN]
button
Executes
span calibration.
10 Calibration alarm state
Displays [Alarm] when a calibration alarm is occurring.
―
Содержание GA-370
Страница 1: ...Trace Gas Monitor GA 370 Instruction Manual eng CODE GZ0000456936B...
Страница 91: ...8 Reference 75 8 5 Related diagrams Name No GA 370 External Dimensions GZ0000471868B Flow Sheet GZ0000465573...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...