![horiba GA-370 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/horiba/ga-370/ga-370_instruction-manual_2162216074.webp)
6 Troubleshooting
58
6
Troubleshooting
6.1
Before troubleshooting
Check the following items again before carrying out troubleshooting.
Is the power switch on?
Is there gas in the cylinder?
Is the main valve of the cylinder open?
Are the span gas concentration and range values correct?
Is the gas flowing normally?
Is a pipe disconnected?
6.2
Failure causes and remedial actions
This section describes trouble centering on the replacement of parts and various check items
that users may perform and the remedial actions to take.
If a problem is not resolved even after taking the remedial actions, be sure to contact HORIBA.
6.2.1 No sensitivity
Cause
Remedial action
Reference
No gas is flowing.
Gas flow rate is insufficient.
Check whether tubes to the analyzer is choked
or disconnected.
―
Apply pressure to the sample gas and
reference gas inlets.
―
Check that the main valve of the span gas
cylinder is open.
―
Check the flow rate.
Check whether the filters are dirty. If they are,
replace them.
Span gas cylinder is empty.
If the residual pressure of the span gas
cylinder is approximately 1 MPa or less,
replace the cylinder.
―
Span gas is incorrect.
Use a span gas that is suitable for the
component to be measured.
―
Gas is leaking.
Replace the joints or otherwise repair the
piping.
―
Others
Contact HORIBA or local representative.
―
Содержание GA-370
Страница 1: ...Trace Gas Monitor GA 370 Instruction Manual eng CODE GZ0000456936B...
Страница 91: ...8 Reference 75 8 5 Related diagrams Name No GA 370 External Dimensions GZ0000471868B Flow Sheet GZ0000465573...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...