4 Various Functions
44
4.4.3 Analog output value check
Whether the current output or voltage output from the analyzer is being performed normally
can be checked.
Operating procedure
1.
Press the button on the right of the channel you wish to output to switch to the
output value.
When the button is pressed, the output values are switched in the following order.
0%
10%
20% to 100%
MEAS
0%
For details on the pin assignment of the connector corresponding to each channel, refer
to "7.1 Analog signal output (ANALOG OUT connector)" (page 65).
Fig. 48 ANALOG OUTPUT screen
2.
Press the analog output button (
).
The value set for each channel is output.
3.
Check the output values.
Note
All channels are set to [MEAS] immediately after the ANALOG OUTPUT screen appears.
[MEAS] indicates the normal state of the output of a measurement value.
When you switch from the ANALOG OUTPUT screen to another screen, the setting of each
channel returns to [MEAS].
DATA MENU
[ANALOG OUTPUT]