3 Operations
32
5.
Select the range to set and then press the [OK] button.
The setting is updated and the RANGE CHANGE screen reappears.
Fig. 32 RANGE CHANGE screen
Table 13 Selection items in RANGE CHANGE screen
Tip
In the case of auto range switching, the range is switched automatically as follows.
Auto range is disable when operating with a calibration range (calibration mode).
When [AUTO] or [EXT] is selected,
([AUTO]) or
([EXT]) is displayed on the right side of
the range status indication in the MEASUREMENT screen (measurement value display).
Item
Description
[1ppm] button
Sets the range to 1 ppm.
[2ppm] button
Sets the range to 2 ppm.
[5ppm] button
Sets the range to 5 ppm.
[10ppm] button
Sets the range to 10 ppm.
[AUTO] button
Sets auto range switching.
The range is switched automatically depending on the measured value.
[EXT] button
Sets external range switching.
The range becomes able to be switched by external contact input.
If the measurement value becomes more than 90% of the current range, the
measurement range switches to the concentration range that is one higher.
If the measurement value becomes less than 80% of the concentration range that is one
lower than the current range, the measurement range switches to the concentration range
that is one lower.
Содержание GA-370
Страница 1: ...Trace Gas Monitor GA 370 Instruction Manual eng CODE GZ0000456936B...
Страница 91: ...8 Reference 75 8 5 Related diagrams Name No GA 370 External Dimensions GZ0000471868B Flow Sheet GZ0000465573...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...