3 Operations
28
3.3.8 Ending calibration
Switch to the delay mode after executing calibration and replace the gas in the analyzer with
the sample gas while remaining in the calibration state.
Operating procedure
1.
Press the [MODE] button in the CALIBRATION screen.
The mode selection screen appears.
2.
Select [DELAY] and then press the [OK] button.
Fig. 27 CALIBRATION screen (during replacement)
3.
When the gas replacement is complete, press the [MODE] button to switch the
mode to [MEASURE].