3.3.6 Canceling calibration alarms
When a calibration alarm is occurring, the calibration error can be canceled by performing this
operation.
Operating procedure
1.
Press the calibration alarm cancel button (
) in the CALIBRATION screen.
Fig. 22 CALIBRATION screen 3 (when calibration alarm occurrence)
2.
A confirmation message appears so press the [YES] button.
The currently occurring calibration alarm is canceled.
Fig. 23 Confirmation message (calibration alarm clear)
Note
When a calibration alarm has occurred, the calibration coefficient is not updated and
measurement is performed using the last calibration coefficient.
The canceling of the calibration alarm is executed simultaneously for all components.