![horiba GA-370 Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/horiba/ga-370/ga-370_instruction-manual_2162216056.webp)
4 Various Functions
40
4.4.1 Periodic replacement parts management
Set the replacement terms and check the current elapsed times of periodic replacement parts.
Furthermore, when the passed time of a periodic replacement part exceeds the recommended
limit time, you can check that replacement is necessary.
For details on periodic replacement parts, refer to Replacement Parts List (page 72).
Check the passed time of each periodic replacement part in the PARTS MANAGEMENT
screen.
If a recommended limit time has been exceeded, the passed time flashes.
Fig. 43 PARTS MANAGEMENT screen
Tip
If a part will not be replaced after the recommended limit time has been exceeded, extend the
recommended limit time or reset the elapsed time.
DATA MENU
[PARTS MANAGEMENT]
Current PASSED TIME
Reset button (passed time is reset to 0)