Annual Inspection
31
EN
8.6
Fluid Deficit Measurement Test
The test setup is depicted in Fig. 8-3, page 31. It is critical that the collection can-
ister be placed
on
the scale as shown in Fig. 8-3.
1. If Basic Function Tests 8.3 to 8.5 have been conducted, skip to step 2. If not,
see Basic Function Tests 8.4 steps 1 to 11.
2. "Zero" the fluid deficit display by pressing the
Zero
button (see Figure 5-1
Item
(9)
).
3. Press the
Pause/Resume
button.
4. Let system run for 1 minute. The canister should have ~800 ml of fluid but the
fluid deficit display should stay at ~0.
5. The acceptable tolerance is ±50.
Record results in the test log in Section 15.1. Test is successful if results fall within
acceptable tolerance limits.
Fig. 8-3
Fluid deficit measurement test
setup
Содержание AQL-100
Страница 6: ......
Страница 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 52: ......
Страница 54: ......
Страница 100: ...48 Anhang DE 15 Anhang 15 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 102: ......
Страница 104: ......
Страница 150: ...48 Annexe FR 15 Annexe 15 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 152: ......
Страница 154: ......
Страница 200: ...48 Apéndice ES 15 Apéndice 15 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 202: ......
Страница 204: ......
Страница 250: ...48 Appendice IT 15 Appendice 15 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 252: ......
Страница 254: ......
Страница 300: ...48 Bijlage NL 15 Bijlage 15 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 302: ......