Safety functions
25
EN
6
Safety functions
The electronic components continuously monitor the proper function of the sys-
tem. System malfunctions are indicated with audible warning tones, error mes-
sages, and/or the blocking of system functions. A table listing a summary of
possible error and warning messages is provided in Chapter 9, Error and Warning
Messages.
Intrauterine pressure 10 mmHg above pre-
set intrauterine pressure setting
If the intrauterine pressure exceeds the intrauterine pressure setting by
10 mmHg for longer than 5 seconds, the pressure reduction function is activated.
The roller wheel will move forward or backward a few times during the pressure
reduction process. If the pressure cannot be reduced, the message
Overpressure/
Open Stopcock
is displayed and three audible tones are emitted.
Intrauterine pressure > 150 mmHg
The message
Maximum Pressure
is displayed and the pump will emit 3 audible
tones once the intrauterine pressure exceeds 150 mmHg. The maximum permis-
sible pressure has now been reached.
Intrauterine pressure > 200 mmHg
If the intrauterine pressure exceeds 200 mmHg for longer than 5 seconds, the
roller wheel stops and the message
Overpressure/Check Stopcock
is displayed.
Three short continuous audible warning tones are continuously emitted until the
pressure is reduced. Once the intrauterine pressure falls below 200 mmHg, the
audible alarm stops and the pump wheel resumes turning automatically.
Check Tube Set Installation
If the inflow tube set is not inserted properly over the roller wheel, pressing the
Prime
button results in a short audible tone and
Check Tube Set Installation
is
displayed. The roller wheel does not start to turn.
Errors of the pressure measuring system
If a malfunction is detected in the pressure measurement electronics,
Sensor Er-
ror
is displayed and five short audible tones are emitted. The roller wheel stops
turning.
Scale overload
If the maximum permissible weight of the scale is exceeded (65 lbs/30 kg), a con-
tinuous audible tone is emitted and
Scale Overloaded Check Scale
is displayed.
The warning stops once the excess weight is removed from the scale.
Loading/unloading scale while in operation
If a container is removed from the scale while the pump is being operated, three
short continuous warning tones are emitted and
Pause Pump for Container
Change
is displayed. The audible warning tone stops once the initial status is re-
stored or the
Pause/Resume
button is pressed
Pressure setting at restart
If the last pre-set intrauterine pressure setting is greater than 80 mmHg, this val-
ue is reset to the default value of 80 mmHg.
Deficit limit
Each additional deficit increase by 100 ml above the selected deficit threshold
triggers 3 audible warning tones that are repeated while the pump continues to
operate. The message
Deficit Limit Exceeded
is displayed.
Deficit rate >300 ml/min
When the deficit rate exceeds 300 ml/min, three audible warning tones are emit-
ted and the message
High Fluid Loss Check Leakage
is displayed. If no obvious
source of high fluid loss can be identified, an assessment of potential cervical or
uterine perforation should be made.
Serious system defect
Five short audible tones are emitted and
Motor Error
is displayed. It is possible
that errors can occur in the start up sequence prior to the enabling of the Pump
Display. In this situation, the Pump Display will remain blank.
Содержание AQL-100
Страница 6: ......
Страница 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 52: ......
Страница 54: ......
Страница 100: ...48 Anhang DE 15 Anhang 15 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 102: ......
Страница 104: ......
Страница 150: ...48 Annexe FR 15 Annexe 15 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 152: ......
Страница 154: ......
Страница 200: ...48 Apéndice ES 15 Apéndice 15 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 202: ......
Страница 204: ......
Страница 250: ...48 Appendice IT 15 Appendice 15 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 252: ......
Страница 254: ......
Страница 300: ...48 Bijlage NL 15 Bijlage 15 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 302: ......