Error and Warning Messages
33
EN
9
Error and Warning Messages
Display messages are indicated visually on the pump display, as well as audibly.
Audible tones to indicate warnings or operation updates (audible tones) are
emitted a certain number of times.
Pump Display Message
Pump Audible Tone
Instructions
Check Tube Set Installation
1 audible tone
Remove and re-insert tube set. If mes-
sage recurs, switch to a new tube set.
Tube Set Over Usage Limits
1 audible tone
Tube set recognition function indicates
tube set has been used already. Insert
new tube set.
Check Flow Path, Stopcock, Clamps
3 audible tones
Flow path is blocked. Check that bag
clamps & hysteroscope stopcock are
open. Check that tube set is free of
obstructions.
Incorrect Tube Set
1 audible tone
Replace tube set. Tube set does not
match type necessary for Aquilex Sys-
tem.
Pause Pump for Container Change
3 audible tones
Removal of canister without activating
Pause button. Insert empty canister and
hit "Pause/Resume" button.
Overpressure Open Stopcock
3 audible tones
Most commonly triggered when hystero-
scope stopcock is closed while pump is
operating at peak flow rate. Open hys-
teroscope stopcock or other occlusion to
relieve pressure.
Overpressure Check Stopcock
5 tones repeatedly continuously until
pressure reduced
Pressure has exceeded 200 mmHg safety
limit and must be reduced. Most proba-
ble cause is hysteroscope being closed
while pump was operating at peak flow
rate.
Open inflow stopcock on hysteroscope or
obstruction that is pinching inflow tube
set.
Maximum Pressure
No audible tone
The message Maximum pressure is dis-
played if the intrauterine pressure
exceeds 150 mmHg.
Deficit Limit Reached
1 audible tone
Physician must respond appropriately.
Deficit Limit Exceeded
3 audible tones
Physician must respond appropriately.
Conduct manual deficit assessment, if
necessary.
Pressure Threshold
1 audible tone
Pressures exceeding 100 mmHg are not
usually required for hysteroscopy. Dili-
gent monitoring of fluid deficit is recom-
mended.
Connect Scale Restart System
3 audible tones
Check scale connection. Reconnect scale
and restart device. If message recurs,
contact Hologic.
Remove Tube Set for System Check
1 audible tone
System check must be performed with
tube set removed from the roller wheel
assembly. Remove tube set and wait for
audible tone and display to indicate
"Insert Tube Set."
Prime Fail -Open Stopcock, Clamps
3 audible tones
Check bag clamp(s) and hysteroscope
inflow stopcock are open. Hit "Prime"
button to restart.
High Fluid Loss Check Leakage
3 audible tones
Physician must respond appropriately.
Conduct manual deficit assessment, if
necessary.
Содержание AQL-100
Страница 6: ......
Страница 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Страница 52: ......
Страница 54: ......
Страница 100: ...48 Anhang DE 15 Anhang 15 1 Testprotokoll Datum Durchgeführte Tests Ergebnis Bemerkung Unterschrift ...
Страница 102: ......
Страница 104: ......
Страница 150: ...48 Annexe FR 15 Annexe 15 1 Procès verbal de tests Date Tests réalisés Résultat Remarque Signature ...
Страница 152: ......
Страница 154: ......
Страница 200: ...48 Apéndice ES 15 Apéndice 15 1 Registro de test Fecha Test realizados Resultado Anotaciones Firma ...
Страница 202: ......
Страница 204: ......
Страница 250: ...48 Appendice IT 15 Appendice 15 1 Protocollo di verifica Data Test eseguiti Risultato Nota Firma ...
Страница 252: ......
Страница 254: ......
Страница 300: ...48 Bijlage NL 15 Bijlage 15 1 Testprotocol Datum Uitgevoerde tests Resultaat Opmerking Handtekening ...
Страница 302: ......