WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
56
Specifications
E
LECTROMAGNETIC
E
MISSIONS
G
UIDANCE
CAUTION:
To help prevent equipment damage, obey these cautions:
•
Caution
—This device meets all requirements for
electromagnetic compatibility per IEC 60601-1-2. It is unlikely
that the user will encounter problems with this device because of
inadequate electromagnetic immunity. However,
electromagnetic immunity is always relative, and standards are
based on anticipated environments of use. If the user observes
unusual device behavior, particularly if such behavior is
intermittent and associated with nearby use of radio or TV
transmitters, cell phones, or electro-surgical equipment, this
could be an indication of electromagnetic-interference. If such
behavior occurs, the user should try to move the interfering
equipment further from this device.
•
Caution
—To safely stop the operation of the system, unplug the
bed and/or the external power supply from the power outlet.
•
Caution
—If the system is installed on a non-Hill-Rom bed (this
system is not available at this time), the system (ME EQUIPMENT)
must be connected to the power outlet (AC mains) by the power
cord supplied with the system. This power cord is semi-
permanently attached to the bed.
WARNING:
Warning
—The P006979 (WatchCare™ reader, indicator light, and
antennas) should not be used adjacent to or stacked with other electrical
equipment. If adjacent or stacked use is necessary, observe the P006979
and the other electrical equipment to make sure they operate as
intended.
Make sure the P006979 operates correctly when it is used near other
electronic devices. Portable and mobile radio frequency (RF)
communications equipment can affect electrical equipment.
Medical equipment needs special precautions in regard to
electromagnetic compatibility (EMC) and needs to be installed and put
DRAFT 26-FEB-2019