WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
48
1.6 Equipment Call Cable Replacement
1.6
Equipment Call Cable Replacement
1.
Make sure the bed is unplugged and you have done the steps for
“Setup” on page 25.
2.
Disconnect the WatchCare™ 1/4" communication cable from the
equipment call jack and from the bed.
3.
Remove the equipment call cable connector from the bed.
VersaCare® Bed
Progressa® Bed
4.
Use the alcohol-based cleaner to clean the area where the equipment
call cable connector was removed. Let the area dry.
CAUTION:
Failure to install the equipment call cable connector in the correct
location could cause equipment damage.
5.
Install the
new
equipment call cable connector as follows:
a.
Remove the adhesive covering from the equipment call cable
connector.
DRAFT 26-FEB-2019