WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
14
Cleaning and Disinfecting
CAUTION:
To help prevent equipment damage, obey these
cautions
:
•
Caution
—Do not steam clean or power wash the system.
Pressure and excessive moisture can damage the protective
surfaces of the system and its electrical components.
•
Caution
—Do not use harsh cleansers/detergents, heavy duty
grease removers, solvents such as toluene, xylene, or acetone,
and do not use scouring pads (you may use a soft bristle brush).
•
Caution
—Do not use bleach as your primary everyday
cleaner/disinfectant.
R
ECOMMENDATIONS
For proper cleaning and disinfection, staff members should be trained.
The
trainer
should carefully read the instructions and follow them when
the
trainee
is being trained. The trainee should:
•
Be given time to read the instructions and to ask any questions.
•
Clean and disinfect the product while the trainer supervises.
During, and/or after this process, the trainer should correct the
trainee of any differences from the instructions for use.
The trainer should supervise the trainee until the trainee can clean and
disinfect the system as instructed.
Hill-Rom recommends to clean and disinfect the system’s components
between patient use and regularly during extended patient stays.
Some fluids used in the hospital environment, such as iodophor and zinc
oxide creams can cause permanent stains. Remove temporary stains by
wiping vigorously with a lightly-dampened wiping cloth.
C
LEANING
AND
D
ISINFECTION
Cleaning and disinfection are distinctly different processes. Cleaning is the
physical removal of visible and non-visible soil and contaminants.
Disinfection is intended to kill microorganisms.
DRAFT 26-FEB-2019