WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
9
Prepare the System for Use
PREPARE THE SYSTEM FOR USE
WARNING:
To help prevent injury and/or equipment damage, obey these
warnings
:
•
Warning
—Keep cables out of the patient foot fall area.
•
Warning
—Connect the WatchCare™ 1/4" communication cable
or WatchCare™ adapter cable (as applicable) to an equipment call
jack designated for the WatchCare™ System only. Patient injury
could occur if the nurse call system is not operational.
1.
Make sure the bed’s power cord is plugged into a power outlet.
NOTE:
Once the bed is plugged in, it may take up to 2 minutes for the indicator
light to turn white.
2.
If the WatchCare™ System interfaces with a nurse call system, make
sure of these:
•
Centrella® Bed
—
–
Bed with the NaviCare® System
—the SideCom® cable is
connected to the bed and to the facility’s equipment call
jack.
–
Bed with a different nurse call system
—the adapter cable
is connected to the bed, and the cable’s 1/4" cable
connector is connected to the facility’s equipment call jack
designated for the WatchCare™ System’s incontinence
alerts.
NOTE:
The location of the adapter cable differs depending on your bed
version. An earlier version bed is shown on the left in the photos
below.
DRAFT 26-FEB-2019