WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
19
Troubleshooting
Progressa® and
VersaCare® Beds
—
the indicator light is
off even though the
LEDs on the reader
flash at power up.
Caregiver
1.
Unplug the bed, and then plug the bed in to
power to reset the bed.
2.
Contact your facility-authorized service
persons.
Service Person Only
1.
Make sure the indicator light is not damaged.
2.
Make sure the indicator light cable is not
damaged and it is fully connected to the
reader and the indicator light P.C. board.
3.
Replace the indicator light assembly (see
page 34).
Progressa® and
VersaCare® Beds
—
one or more of the
indicator light col-
ors do not come on.
Caregiver
1.
Unplug the bed, and then plug the bed in to
power to reset the bed.
2.
Contact your facility-authorized service
persons.
Service Person Only
1.
Make sure the indicator light is not damaged.
2.
Make sure the indicator light cable is not
damaged and it is fully connected to the
reader and the indicator light P.C. board.
3.
Replace the indicator light assembly (see
page 34).
Problem
Solution
DRAFT 26-FEB-2019