WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
38
1.3 Indicator Light Assembly Replacement
2.
Follow the routing of the old light assembly cable to the reader, and
then use the same routing for the
new
light assembly cable. Remove
and install cable ties as necessary.
NOTE:
At the head end of the guide
tube, you may need to use a
screwdriver to move the plastic
cover enough to get access to
the old and new cables.
•
After you route the new cable through the guide tube, do these:
a.
Pull the old cable out from the guide tube.
CAUTION:
Caution
—The foot section must be extended to correctly install the
indicator light. Otherwise, equipment damage could occur.
b.
Extend the foot section.
c.
Make sure that the footboard sensor switch is fully
connected, and then install the sliding foot cover (D) and its
fourteen screws (C) on to the bed (see Step 2 on page 36).
d.
Continue routing the new cable to the reader.
DRAFT 26-FEB-2019