2.8 Hilow Actuator Malfunction
Chapter 2: Troubleshooting Procedures
RESIDENT® Low Bed Service Manual (man259)
Page 2 - 37
2
Yes
No
↓
→
If a only one button produces a click when pushed, go to RAP
2.5. If there is no click when either button is pushed, replace the
electrical control/lockout box (refer to procedure 4.9).
4. Check the hilow actuator motor cable connection at the electrical
control/lockout box.
The connection is sound.
Yes
No
↓
→
Insert the loose plug firmly into its socket. If this solves the
problem, go to “Final Actions” on page 2-16. If not, go to step 5.
5. Check the condition of the cable between the electrical control/lockout box
and the hilow actuator.
The cable shows no signs of damage.
Yes
No
↓
→
If the cable is damaged, replace the hilow actuator (refer to
procedure 4.12).
6. Press and hold the hilow up button (A) on the hand pendant.
The bed rises smoothly until it is between 26" and 26½" (66 cm and 67.31
cm) above the floor, and it stops automatically.
Yes
No
↓
→
Check for interference from other parts of the bed by pressing
and holding the hilow down button (B) on the hand pendant If
interference is found, correct it. If no interference is found,
replace the hilow actuator (refer to procedure 4.12).
7. Press and hold the hilow down button (B) on the hand pendant.
The bed lowers smoothly until it is between 10½" and 11" (26.67 cm and
28 cm), and it stops automatically.
Yes
No
↓
→
Check for interference from other parts of the bed. If interference
is found, correct it. If no interference is found, replace the hilow
actuator (refer to procedure 4.12).
8. Press the hilow up (A) button on the hand pendant, and raise the bed until it
is approximately 20" (50.8 cm) above the floor.
At the electrical control/lockout box, push in the hilow actuator lockout button
(C) (see figure 2-17 on page 2-38).
Содержание RESIDENT P872
Страница 2: ......
Страница 6: ...Revisions Page iv RESIDENT Low Bed Service Manual man259 NOTES...
Страница 12: ...Table of Contents Page x RESIDENT Low Bed Service Manual man259 NOTES...
Страница 32: ...Chapter 2 Troubleshooting Procedures Page 2 2 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 78: ...Chapter 3 Theory of Operation Page 3 2 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 88: ...Theory of Operation Chapter 3 Theory of Operation Page 3 12 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 136: ...Chapter 5 Parts List Page 5 2 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 138: ...Warranty Chapter 5 Parts List Page 5 4 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 143: ...Recommended Spare Parts Chapter 5 Parts List RESIDENT Low Bed Service Manual man259 Page 5 9 5 NOTES...
Страница 149: ...Main Frame Chapter 5 Parts List RESIDENT Low Bed Service Manual man259 Page 5 15 5 NOTES...
Страница 155: ...Siderail Assemblies Chapter 5 Parts List RESIDENT Low Bed Service Manual man259 Page 5 21 5 NOTES...
Страница 156: ...Labels Chapter 5 Parts List Page 5 22 RESIDENT Low Bed Service Manual man259 Labels Figure 5 6 Labels m259_003...
Страница 158: ...Labels Chapter 5 Parts List Page 5 24 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 160: ...Chapter 6 General Procedures Page 6 2 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 168: ...Tool and Supply Requirements Chapter 6 General Procedures Page 6 10 RESIDENT Low Bed Service Manual man259 NOTES...
Страница 170: ...Chapter 7 Accessories Page 7 2 RESIDENT Low Bed Service Manual man259 NOTES...