Function Checks—Electric Drive Model
Chapter 2: Troubleshooting Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 2 - 13
2
Figure 2-6. Caregiver Siderail Control Panel (Hilow Actuator Controls)
11. Press the HEAD Up button (A) on the handset control panel to raise the
head section to its greatest inclined position (see figure 2-7 on page 2-13).
The head of the bed raises fully and evenly to approximately 65
o
as
indicated by the head angle indicator decals located on both sides of the
bed.
Yes
No
↓
→
Figure 2-7. Handset Control (Head Actuator Controls)
12. Press the HEAD Down button (B) to lower the head section fully. The head
section of the bed lowers fully and evenly to a flat position.
Yes
No
↓
→
B
A
m136_013
m136_012
A
B
Содержание P870
Страница 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Страница 51: ...Warning and Caution Labels Chapter 1 Introduction Page 1 38 ResidentTM LTC Bed Service Manual man136rc 1 NOTES...
Страница 53: ...Chapter 2 Troubleshooting Procedures Page 2 2 ResidentTM LTC Bed Service Manual man136rc 2 NOTES...
Страница 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Страница 111: ...Chapter 4 Removal Replacement and Adjustment Procedures Page 4 4 ResidentTM LTC Bed Service Manual man136rc 4 NOTES...
Страница 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Страница 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Страница 170: ...Recommended Spare Parts Chapter 5 Parts List ResidentTM LTC Bed Service Manual man136rc Page 5 9 5 NOTES...
Страница 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...