360077-1/O
109
Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
23.4
Move Slides and Brackets
Remove drawer slides
1
Using a screwdriver, remove front bracket retainers.
2
Tap front brackets upward to disengage standards.
3
Remove slides from standards.
Install drawer slides
1
Insert slides into standard at appropriate height.
2
Tap front brackets downward to engage standards.
3
Using a screwdriver, install front bracket retainers.
Remove shelf brackets
1
Using a screwdriver, remove front bracket retainers.
2
Tap front brackets upward to disengage standards.
3
Remove front brackets from standards.
Install shelf brackets
1
Insert front brackets into standard at appropriate height.
2
Tap front brackets downward to engage standards.
3
Using a screwdriver, install front bracket retainers.
23.5
Optional Adapter Kits for Medication Dispensing Locks
Contact Helmer Technical Service or your distributor for service documentation pertaining to medication
dispensing locks.
24 Settings
Through the Laboratory combined monitor and controller, current settings may be viewed and changed.
NOTICE
►
Control Sensor Offset and Hysteresis settings are factory-preset and should not be
changed unless directed by Helmer Technical Service.
►
Changing temperature settings affects operation of the refrigerator. Do not change
settings unless instructed by Helmer Technical Service.