360077-1/O
111
Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
24.2
Refrigerator Setpoint
NOTE
Default setpoint is 4.0 °C.
Change the setpoint if:
►
Your organization requires a chamber temperature other than 4.0 °C.
►
The normal chamber temperature is too high or low (after completing preventive maintenance and
applicable troubleshooting tasks).
Confirm:
►
Refrigerator has been placed per location requirements in the operation manual.
►
Preventive maintenance has been completed per operation manual.
►
Troubleshooting items associated with chamber temperature have been reviewed (if necessary).
Change setpoint.
1
Determine the change in value to reach desired setpoint. Adjustment should be the difference
between current setpoint and new setpoint.
EXAMPLE
►
Current setpoint is 4.0 °C
►
Target setpoint is 5.0 °C
►
Setpoint adjustment value is +1.0 °C
2
On the monitoring system, press and release
SEL
to change to Control mode. CONTROL lamp will
illuminate.
3
Press and hold
SET
to display the reference temperature.
4
Hold
SET
and press
Up Arrow
or
Down Arrow
as necessary to set the adjustment value determined
in step 2.
5
Release all buttons; the setpoint is changed.
6
Press and release
SEL
to return to Monitor mode. MONITOR lamp will illuminate.
24.3
Temperature Alarm Setpoints
View setpoints:
1
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp will flash to indicate entry into
program mode.
2
Press
SEL
until desired setting appears.
3
Observe the setting.
4
To view another setting, press
SEL
again (optional).
5
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp stops flashing to indicate exit
from program mode.
Flashing Lamp
Selected Setting
HIGH TEMP and MONITOR
High temp alarm setpoint
LOW TEMP and MONITOR
Low temp alarm setpoint
MONITOR only
Monitor offset
CONTROL only
Control sensor offset
CONTROL only
Control hysteresis