Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
24.3.1
High Temperature Alarm
►
Specifies the temperature at which the High Temperature Alarm activates.
►
Default setpoint is 5.5 °C.
►
Setpoint can be changed from -40.0 °C to +25.0 °C.
Change the setpoint:
1
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp will flash to indicate entry into
program mode.
2
Press
SEL
until HIGH TEMP and MONITOR lamps flash.
3
Hold
SET
, then press
Up Arrow
or
Down Arrow
to change the setpoint.
4
Release
SET
button.
5
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp stops flashing to indicate exit
from program mode. New settings are saved.
24.3.2
Low Temperature Alarm
►
Specifies the temperature at which the Low Temperature Alarm activates.
►
Default setpoint for HHB (blood bank) models is 1.5 °C.
►
Default setpoint for HLR and HPR (laboratory and pharmacy) models is 2.0 °C.
►
Setpoint can be changed from -40.0 °C to +25.0 °C.
Change the setpoint:
1
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp will flash to indicate entry into
program mode.
2
Press
SEL
until LOW TEMP and MONITOR lamps flash.
3
Hold
SET
, then press
Up Arrow
or
Down Arrow
to change the setpoint.
4
Release
SET
button.
5
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp stops flashing to indicate exit
from program mode. New settings are saved.
24.4
Temperature Calibration Setpoints
View setpoints:
1
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp will flash to indicate entry into
program mode.
2
Press
SEL
until desired setting appears.
3
Observe the setting.
4
To view another setting, press
SEL
again (optional).
5
Hold
Up Arrow
and
Down Arrow
for three seconds. MONITOR lamp stops flashing to indicate exit
from program mode.
Flashing Lamp
Selected Setting
HIGH TEMP and MONITOR
High temp alarm setpoint
LOW TEMP and MONITOR
Low temp alarm setpoint
MONITOR only
Monitor offset
CONTROL only
Control sensor offset
CONTROL only
Control hysteresis