360077-1/O
32
i.Series® - All Models
8.9.3
Power Failure Alarm Test
NOTE
►
During a power failure, the power failure alarm activates and the batteries provide
power to the monitoring system.
►
If AC power fails, the backup batteries will allow for continued data collection and
temperature display.
►
If the backup batteries fail, data is not collected and the temperature is not displayed.
►
When power is restored, the 24 hours of data prior to the power loss are retained
in the system memory. Stored temperature data is displayed on the graph and the
monitoring system resumes data collection and display.
1
Confirm the refrigerator is connected to AC power.
2
Ensure the monitoring system backup batteries are installed.
3
Change Power Failure Timeout setting to 0 minutes.
a
Press the
MAIN
button.
b
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
c
Enter the password when prompted.
d
Press the
DOWN
button to highlight Alarm Setpoints. Press the
SELECT
button.
e
Press the
DOWN
button to highlight Power Failure Timeout.
f
Press the
DEC
button to change the setting to 0.
4
Disconnect the refrigerator from AC power. Power failure alarm will activate immediately.
5
Reconnect the refrigerator to AC power. Power failure alarm will clear and audible alarm will cease.
6
Change the Power Failure Timeout setting to the original setting.
8.9.4
Door Ajar Alarm Test
1
Change Door Ajar Timeout setting to 0 minutes:
a
Press the
MAIN
button.
b
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
c
Enter the password when prompted.
d
Press the
DOWN
button to highlight Alarm Setpoints. Press the
SELECT
button.
e
Press the
DOWN
button to highlight Door Ajar Timeout.
f
Press the
DEC
button to change the setting to 0.
g
Press the
BACK
button to return to the Edit Configuration screen, or press the
HOME
button to
exit. New settings are saved.
2
Open the door. Door ajar alarm will activate immediately.
3
Close the door. Door ajar alarm will clear and audible alarm will cease.
4
Change the Door Ajar Timeout setting to the original setting.
8.9.5
No Battery Alarm Test
Test the no battery alarm to ensure the alarm provides warning of a low- or no-battery charge.
NOTE
►
During an AC power failure, the power failure alarm activates and the batteries
provide power to the monitoring system.
►
If AC power fails, the backup batteries will allow for continued data collection and
temperature display.
►
If the backup batteries fail, data is not collected and the temperature is not displayed.