360077-1/O
78
Horizon Series™ - Blood Bank Models
16.8.2
Power Failure Alarm Test
NOTE
►
During a power failure, the power failure alarm activates and the batteries provide
power to the monitoring system.
►
If AC power fails, the backup batteries will allow for continued temperature display.
►
If the backup batteries fail, the temperature is not displayed.
►
When power is restored, the monitoring system resumes data display.
1
Confirm the refrigerator is connected to AC power.
2
Ensure the monitoring system backup batteries are installed.
3
Change Power Failure Timeout setting to 0 minutes.
a
Press the
MAIN
button.
b
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
c
Enter the password when prompted.
d
Press the
DOWN
button to highlight Alarm Setpoints. Press the
SELECT
button.
e
Press the
DOWN
button to highlight Power Failure Timeout.
f
Press the
DEC
button to change the setting to 0.
4
Disconnect the refrigerator from AC power. Power failure alarm will activate immediately.
5
Reconnect the refrigerator to AC power. Power failure alarm will clear and audible alarm will cease.
6
Change the Power Failure Timeout setting to the original setting.
16.8.3
Door Ajar Alarm Test
1
Change Door Ajar Timeout setting to 0 minutes:
a
Press the
MAIN
button.
b
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
c
Enter the password when prompted.
d
Press the
DOWN
button to highlight Alarm Setpoints. Press the
SELECT
button.
e
Press the
DOWN
button to highlight Door Ajar Timeout.
f
Press the
DEC
button to change the setting to 0.
g
Press the
BACK
button to return to the Edit Configuration screen, or press the
HOME
button to
exit. New settings are saved.
2
Open the door. Door ajar alarm will activate immediately.
3
Close the door. Door ajar alarm will clear and audible alarm will cease.
4
Change the Door Ajar Timeout setting to the original setting.