Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
Section IV: Horizon Series™ - Laboratory, Pharmacy, and International
Blood Bank Models
NOTE
This section applies to HHB, HLR, and HPR models.
23
Product Configuration
23.1
Install Battery for Backup Power
The monitoring system has a battery system, enabling a period of continuous operation if power is lost.
NOTE
The monitoring system will start on battery power alone. If the refrigerator was previously
not connected to AC power and the battery is connected, the monitoring system will begin
running on battery power.
Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available
and no battery-related alarms are active, backup power for the monitoring system is available for up to
two hours.
NOTICE
When installing replacement batteries, use only batteries which meets the specifications
outlined in chapter
25.7
(Supplies).
The battery is located on the top of the refrigerator, behind the monitoring system. For 111 models, a
service cover covers the components and an access panel provides access to the monitoring system
backup battery.
Monitoring system backup battery.
Monitoring system battery is included in the accessory package. Install and connect the battery to
provide monitoring system with backup power in the event of an AC power failure.