360179-A/B
48
Troubleshooting
Problem
Possible Cause
Action
The chamber
temperature does
not stabilize at the
Compartmental
Access Refrigerator
setpoint.
Upper chamber probe
bottle is not at proper
solution level
►
Refill bottle if necessary
(refer to
Section II, Item
7 .1 .2
).
Ambient air
temperature around the
Compartmental Access
Refrigerator is too warm
►
Confirm the Compartmental Access Refrigerator is
placed appropriately (refer to the
Section II, Item 5
for location requirements).
Air circulation at the
top of the chamber is
not adequate
►
Ensure that the guard panel and recirculation
panel are installed between the unit cooler and the
compartment assembly.
►
Ensure no obstructions have been placed on the
guard panel.
Condenser grill is dirty
►
Check the condenser grill. Clean the grill if
necessary (refer to
Section III, Item 9 .6 .1)
.
Condenser fan is not
running
►
Check the condenser fan wiring connection.
Secure the connection if necessary.
►
If the wiring connection is secure, replace the fan
motor.
Unit cooler fan is not
running
►
Check for voltage to the fan when door switch is
closed at CP board J44 NO pin.
►
If voltage is present when switch is closed, replace
the fan motor.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Refrigerant level is too
low
►
Check the refrigeration lines for leaks and repair
them if necessary. Check the refrigerant level.
Recharge the refrigerant if necessary.
Compressor solid
state relay is faulty
►
Confirm the control board indicates the
compressor should be running. Touch i.C³ APPS,
Settings (enter the Settings password), Device
Status.
►
If the compressor should not be running, check the
compressor solid state relay.
►
If the compressor solid state relay is closed,
replace the relay.
Compressor motor
has seized
►
Replace the compressor.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
A component is
faulty or internal
connections are loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.