360179-A/B
14
Installation & Configuration
7
Attach waste basket to Consumable Cart.
Attach waste basket to cart
8
Close and lock the door.
6 .3 .3
Attach the Access Console to a Wall
A wall mounting kit is required to attach the Access Console to the wall. Refer to the instructions included
with the wall mounting kit for installation information. Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support. Part number 401323-1.
6 .3 .4
Attach the Access Console to a Table
A table mounting kit is required to attach the Access Console to a table Refer to the instructions included
with the table mounting kit for installation information. Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support. Part number 401322-1.
6 .4
Connect External Monitoring Devices
CAUTION
►
The interface on the remote alarm monitoring system is intended for connection to
the end user’s central alarm system(s) that uses normally-open or normally-closed
dry contacts.
►
If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the
remote alarm monitoring system’s circuit, the remote alarm will not function properly;
may be damaged; or may result in injury to the user.
Required tools:
►
#2 Phillips screwdriver
►
Cable tie
The remote alarm interface is a relay switch with three terminals:
►
Common (COM)
►
Normally Open (NO)
►
Normally Closed (NC)
The terminals on the remote alarm interface have the following maximum load capacity:
►
0.5 A at 30 V (RMS); 1.0 A at 60 V (DC)
Connect to the alarm interface:
1
On the electrical box, locate the remote alarm terminals.
►
Terminals are located on the electrical box, on top of the Compartmental Access Refrigerator