360179-A/B
28
Maintenance & Service
9.1.3
Power Failure Alarm Test
1
Change the Power Failure delay setting to 0 minutes.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
–
on the Power Failure spin box to change the value to 0.
2
Switch the AC ON/OFF switch
OFF
.
►
Power Failure alarm will activate immediately
3
Switch the AC ON/OFF switch
ON
.
►
Power Failure alarm will clear and audible alarm will cease
4
Change the Power Failure time delay to the original setting.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
+
on the Power Failure spin box to change the value to the original setting.
5
Touch
Home
.
9 .1 .4
Door Open Alarm Test
1
Change the Door Open (Time) delay setting to 0 minutes.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
–
on the Door Open (Time) spin box to change the value to 0.
2
Open the door.
►
Door Open alarm will activate immediately
3
Close the door.
►
Door Open alarm will clear and audible alarm will cease
4
Change the Door Open (Time) setting to the original setting.
a
Touch
Home
,
i.C³ APPS
,
Settings
.
b
Enter the Settings password.
c
Touch
Alarm Settings
.
d
Touch
+
on the Door Open (Time) spin box to change the value to the original setting.
5
Touch
Home
.
9.2
Upgrade System Firmware
Helmer may occasionally issue updates for the i.C³ firmware. Follow the upgrade instructions included
with the firmware update.
9 .3
Test and Replace Backup Batteries
9 .3 .1
Test the i .C
3
®
Monitoring System / Access Control Backup Battery
Required tools:
►
4’ ladder or step stool