360179-A/B
20
Installation & Configuration
7 .2 .1
Chart Recorder Access
Open door by pressing and releasing the door.
Chart recorder door
7.2.2
Install Chart Paper
1
Press and hold
C
button. When stylus begins to move left, release button. The LED flashes.
2
When stylus stops moving, remove chart knob then move knob up and away.
3
Place chart paper on chart recorder.
4
Gently lift stylus and rotate paper so current time line corresponds to time line groove.
Chart recorder stylus and time line groove
5
Hold chart paper and reinstall chart knob.
NOTE
For accurate temperature reading, ensure that the current time is aligned with time line
groove when chart knob is tightened.
6
Press and hold
C
button. When stylus begins to move right, release button.
7
Confirm stylus is marking temperature correctly.
7.2.3
Reorder Chart Paper
Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support. Chart paper: Part number 220366
(52 sheets)