360179-A/B
32
Maintenance & Service
9 .5
Inspect and Secure Electrical Wiring and Terminals
WARNING
Disconnect HaemoBank™ from AC power when inspecting and securing wiring terminals.
Required tools:
►
#2 Phillips screwdriver
►
small blade screwdriver
►
4’ ladder or step stool
Inspect components and secure terminals:
1
Switch the Compartmental Access Refrigerator and Access Console battery ON/OFF switches
OFF
.
Switch the AC ON/OFF switch
OFF
. Disconnect the power cord from AC power.
2
Using a #2 Phillips screwdriver, loosen the screws securing the cover to the electrical box.
3
Lift the cover upward and remove from the electrical box.
4
Visually inspect the wiring in the electrical box for discoloration.
►
If any discoloration is found, contact Haemonetics
®
Corporation BloodTrack
®
Customer Support
Terminal Strips
5
Gently pull on all wires which are terminated with a connector.
►
If any wires are loose in the terminal connector, replace the connector
►
If any terminal connectors are loose in the terminal strip(s), use a small blade screwdriver to
tighten the screw in the terminal strip
6
Replace the cover on the electrical box.
7
Using a #2 Phillips screwdriver, tighten the screws securing the cover.
8
Reconnect the power cord to AC power. Switch the AC ON/OFF switch
ON
. Switch the
Compartmental Access Refrigerator and Access Console battery ON/OFF switch
ON
.