Maintenance & Service
Alarm Settings
Restored Value
Upper Probe Low Temperature (Setpoint)
2.0 ºC
Upper Probe Low Temperature (Time Delay)
0 minutes
Air Probe High Temperature (Setpoint)
14.0 ºC
Air Probe High Temperature (Time Delay)
3 minutes
Air Probe Low Temperature (Setpoint)
-6.0 ºC
Air Probe Low Temperature (Time Delay)
3 minutes
Compressor Temperature (Setpoint)
50.0 ºC
Compressor Temperature (Time Delay)
5 minutes
Power Failure
1 minute
Probe Failure
0 minutes
Door Open (Time)
10 minutes
8 .6
Restore Factory Default Settings
NOTICE
►
If Factory Default Settings are restored, all calibration values and User PINs will be
lost.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support prior to restoring
Factory Default Settings.
Restore settings:
1
Touch
Home
,
i.C³ APPS
,
Settings
.
2
Enter the Settings password.
3
Touch
Restore Factory Settings
.
►
The
“Are you sure you want to restore factory settings?”
message is displayed
►
Touch
√
. The message screen closes and factory settings are restored
►
Touch
X
. The message screen closes and factory settings are not restored
4
Touch
Home
.
8.7
View System Information
1
Touch
Home
,
i.C³ APPS
,
Contacts
.
►
The Contacts screen is displayed
2
Touch
More Info
.
►
The Versions screen is displayed, including software configuration, software revision numbers,
hardware serial numbers, i.C³ IP address, and i.C³ MAC address
3
Touch
Home
.