360179-A/B
23
Installation & Configuration
8 .3 .1 .1
Temperature Setpoint
NOTE
►
Default chamber temperature setpoint is 4.0 ºC.
►
Setpoint can be changed from the i.C
3
and from the Haemonetics
®
BloodTrack
®
interface.
The temperature setpoint is the desired temperature of the HaemoBank™ for stored product.
Change the temperature setpoint if:
►
Your organization requires a chamber temperature other than 4.0 ºC
Perform the following:
1
Touch
Home
,
i.C³ APPS
,
Settings
.
2
Enter the Settings password.
3
Touch
+
or
–
on the
Temperature Setpoint
spin box to select desired setting.
4
Touch
Home
.
8 .4
Sensor Calibration
Sensor calibration values are programmed at the factory. Calibration values can be viewed through the
i.C³ Sensor Calibration screen.
Sensor Calibration screens
8.4.1
View Sensor Calibration Values
1
Touch
Home
,
i.C³ APPS
,
Settings
.
2
Enter the Settings password.
3
Touch
Sensor Calibration
.
►
Sensor offset values and their current temperature readings are displayed
4
Touch
Home
.
Setting
Initial Factory Value
Simulated product
probe
Varies (set at the factory to match
a NIST traceable independent
thermometer)
Control probe
Air probe
Compressor probe