360179-A/B
29
Maintenance & Service
Test the battery:
1
Switch the AC ON/OFF switch
OFF
.
►
i.C³ screen should continue to display information with reduced brightness
►
The battery icon will appear on the screen
►
If the Low Battery alarm activates, or the display is blank, replace the battery
2
Switch the AC ON/OFF switch
ON
.
9 .3 .2
Replace the i .C
3
®
Monitoring System / Access Control Backup Battery
NOTICE
Replacement battery must meet the following specifications: 12V, 7Ah, rechargeable,
sealed, lead acid. Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support for
replacement battery. Service Kit Part number 800251-1.
Required tools:
►
#2 Phillips screwdriver
►
4’ ladder or step stool
Replace the battery:
1
Switch the i.C³ battery ON/OFF switch
OFF
. Switch the AC ON/OFF switch
OFF
.
2
Disconnect the power wires from the battery terminals.
3
Using a #2 Phillips screwdriver, remove the screws securing the battery strap to the top of the
compartment.
4
Remove the strap and set it aside.
5
Remove the battery from the HaemoBank™.
6
Place the new battery on the top of the compartment, in the same orientation as the original battery.
7
Place the strap over the battery.
8
Using a #2 Phillips screwdriver, install the screws to attach the strap to the top of the compartment.
9
Connect the power wires to the battery terminals, ensuring the wires are connected to the correct
terminals:
►
Red (+) wire connected to the red (+) battery terminal
►
Black (-) wire connected to the black (-) battery terminal
10
Switch the AC ON/OFF switch
ON
. Switch the i.C³ battery ON/OFF switch
ON
.
9 .3 .3
Check the Access Console Backup Battery
Check the battery:
1
With the BloodTrack
®
kiosk operational, switch the HaemoBank™ AC ON/OFF switch
OFF .
►
BloodTrack
®
kiosk should continue to display information
►
If the display is blank, replace the battery
2
Switch the HaemoBank™ AC ON/OFF switch
ON
.
9 .3 .4
Replace the Access Console Backup Battery
NOTICE
Replacement battery must meet the following specifications: 12V, 7Ah, rechargeable,
sealed, lead acid. Contact Haemonetics
®
Corporation BloodTrack
®
Customer Support for
replacement battery. Service Kit Part number 800215-1.
Required tools:
►
#2 Phillips screwdriver
►
4’ ladder or step stool