![HBK DSE Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/hbk/dse/dse_operating-manual_2111691057.webp)
55
DSE
WEB INTERFACE
Filter settings
Graphical representation
Measured value position mark
ers
Fig. 7.10
Filter menu
Position markers
There are two position markers at the left and right margins that you can drag over the
measurement values using your mouse. The exact measurement values and the
difference between the two positions are displayed directly below the graph. The
difference is additionally converted into a frequency Click in the area between the
markers to move this range.
Your load cell readings may be distorted by external vibration (motors, plant vibrations,
etc.). You should first try to determine the frequency of the vibration so as to filter it out
in a targeted manner. Use the two position markers and the frequency display to do this.
For easier distinction, the signals shown (Unfiltered/Device setting/Simulation) are
displayed in different colors. This allows you to see the effect of filter settings directly in
the graph. As long as you do not apply changes, the changes will be displayed in the
simulation mode; after applying them, they will be displayed in the device setting.
Unfiltered is the original signal at the input of the filter stages. Use
in the graph to
show or hide signals.
Содержание DSE
Страница 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Страница 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Страница 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Страница 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Страница 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Страница 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Страница 203: ...DSE ENTSORGUNG 102 ...
Страница 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...