WEB INTERFACE
Tip
Select the parameter set in which you want to save the settings before making an
adjustment. If you select a parameter set afterward, the settings for that parameter set
will be loaded, overwriting your own.
Automatic calibration
The DSE can perform the adjustment by measuring with an exact weight (test or
calibration weight):
u
In the Calibration area click on Automatic.
u
Click on
RECALIBRATION
.
u
Enter the desired Measuring unit and Rated capacity of your load cell.
u
Click on
CONTINUE
.
u
Remove any weight from the scale and click ADJUST. This measures the zero point.
u
Click on CONTINUE.
u
Enter the test weight and click on
to save it.
u
Load the scale with the test weight (calibration weight).
u
Click on ADJUST and wait for the measurement to finish.
u
Click on CONTINUE.
Finally, the measurement results of the adjustment are displayed. Click on
RECALIBRATION
to start a new adjustment procedure.
Содержание DSE
Страница 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Страница 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Страница 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Страница 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Страница 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Страница 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Страница 203: ...DSE ENTSORGUNG 102 ...
Страница 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...