51
DSE
WEB INTERFACE
Manual adjustment
Perform a manual adjustment as follows:
u
In the Calibration area click on Manual.
u
Enter the desired Measuring unit and Rated capacity of your load cell.
u
Enter the values for the zero load at x1: Value unloaded in mV/V and for rated
capacity at x2: Value at rated capacity in mV/V. The values must be entered in the
unit mV/V. If you have values for a specific bridge excitation voltage, you must
convert them to 1V.
u
Confirm each entry by clicking on
.
Important
The results of the adjustment are initially only stored temporarily. You can save them
permanently in the device, like all value changes, in the Parameter Sets menu.
Fig. 7.8
Manual adjustment