2

 DEVICE DESCRIPTION 

18 

Senator S24S

01/2019 - Senator S24S Radar Manual v1.00 en

2.2  Device description

This device is a 24 GHz FMCW-radar level transmitter. It is a non-contact technology and is 
2-wire loop-powered. It is designed to measure the distance, level, mass, volume and reflectivity
of granulates and powders. For more data about the measuring principle, refer to 

Measuring 

principle

 on page 123.

Radar level transmitters use an antenna to emit a signal to the surface of the measured product. 
The device has many antennas available. Thus, it can measure most products even in difficult 
conditions. Also refer to 

Technical data

 on page 123.

If the device is ordered with the applicable options, it can be certified for use in hazardous areas.

The signal converter is attached directly to the process connection and the antenna. The 
illustration that follows shows the types of antenna.

Figure 2-2: Types of antenna

1  

Metallic Horn antennas (available antenna sizes: DN80 (3

¨

), DN100 (4

¨

), DN150 (6

¨

) and DN200 (8

¨

)) made of 316L stain-

less steel. Antenna extensions are available for high nozzles.

2  

Drop antennas (available antenna sizes: DN80 (3

¨

) DN100 (4

¨

) and DN150 (6

¨

)) made of PTFE or PP. Antenna extensions 

are available for high nozzles.

Accessories: 

Accessories: 

Accessories: 

Accessories: 
For more data about general items, refer to Accessories: general items on page 166

. For more 

data about process connections, refer to 

Accessories: process connection kits on page 167

.

www.hawkmeasurement.com

Содержание Senator S24S

Страница 1: ...Hz Radar FMCW Level Transmitter for solids from granulates to rocks Senator S24S Senator S24S Senator S24S Senator S24S Handbook Handbook Handbook Handbook HAWK 01 2019 Senator S24S Radar Manual v1 00...

Страница 2: ...uce this documentation or any part thereof without the prior written authorisation of Hawk Measurement Systems Pty Ltd 2 www hawkmeasurement com 01 2019 Senator S24S Radar Manual v1 00 en Copyright 20...

Страница 3: ...ate examples 20 3 Installation 21 3 1 General notes on installation 21 3 2 Storage 21 3 3 Transport 22 3 4 Pre installation requirements 22 3 5 Pressure and temperature ranges 23 3 6 Recommended mount...

Страница 4: ...al mode 60 6 3 Program mode 64 6 3 1 General notes 64 6 3 2 Protection of the device settings access levels 65 6 3 3 How to get access to the Quick Setup menu 67 6 3 4 Keypad functions 68 6 3 5 How to...

Страница 5: ...description 150 9 2 Software history 151 9 3 Connection variants 151 9 3 1 Point to Point connection analogue digital mode 151 9 3 2 Multi Drop connection 2 wire connection 152 9 4 HART device variabl...

Страница 6: ...and effect on compatibility 1 Downwards compatible changes and fault repair with no effect on operation e g spelling mistakes on display 2 _ Downwards compatible hardware and or software change of int...

Страница 7: ...more data about the EU Directives and European Standards related to this device refer to the EU Declaration of Conformity You can download this document free of charge from the website All devices ar...

Страница 8: ...ar TLPR Tank Level Probing Radar TLPR Tank Level Probing Radar devices measure level in a closed space only You can use LPR devices for TLPR applications The LPR and TLPR devices meet the requirements...

Страница 9: ...nk shall be closed to ensure a low level leakage of the signal into the air outside the tank whenever possible mounting of the TLPR equipment shall be on top of the tank structure with the orientation...

Страница 10: ...st get the approval of the national regulatory authority before installation e g ANFR France Bundesnetzagentur Germany Ofcom United Kingdom etc Country Name of the station Location Latitude Longitude...

Страница 11: ...ies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received includin...

Страница 12: ...damage this device However devices found to interfere with primary licensing operations will be required to be removed at the user s expense 3 The TLPR device shall be installed and operated in a com...

Страница 13: ...on Identification Number This number gives the radar signal frequency 24GHZ 24 GHz the location of the device T TLPR or L LPR and the type of signal converter compact C TLPR device HVIN 24GHZ T C LPR...

Страница 14: ...g when communicating by e mail may involve gaps in security It is not possible to protect such data completely against access by third parties We hereby expressly prohibit the use of the contact data...

Страница 15: ...ury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards safety requirements and accident prevention regulations...

Страница 16: ...e observed without fail Even partial disregard of this warning can lead to serious health problems and even death There is also the risk of seriously damaging the device or parts of the operator s pla...

Страница 17: ...antenna extensions are too long the de vice will be supplied in two parts 3 Quick Start and supplementary instructions if the device has the appropriate options 4 Bar magnet 5 Display extractor for re...

Страница 18: ...page 123 If the device is ordered with the applicable options it can be certified for use in hazardous areas The signal converter is attached directly to the process connection and the antenna The il...

Страница 19: ...2 3 Visual check 1 Device nameplate for more data refer to Nameplate examples on page 20 2 Process connection data size and pressure rating material reference and heat number 3 Gasket material data r...

Страница 20: ...tage and maximum current fieldbus options basic current 4 Degree of ingress protection according to EN 60529 IEC 60529 5 6 Date of manufacture 7 Serial number 8 Type code defined in order For more dat...

Страница 21: ...list to make sure that you have all the elements given in the order Look at the device nameplate to ensure that the device is delivered according to your order Check for the correct supply voltage pr...

Страница 22: ...signal converter to heavy vibrations The devices are tested for vibration and agree with EN 50178 and IEC 60068 2 6 Figure 3 2 How to lift the device WARNING Lift the device carefully to prevent damag...

Страница 23: ...le that follows Devices with Hazardous Location approvals see supplementary instructions 2 Ambient temperature for operation of the display 20 70 C 4 158 F If the ambient temperature is not between th...

Страница 24: ...lids 1 Position of the process fitting from the silo wall r 2 for DN80 DN100 DN150 or DN200 horn antennas and DN80 DN100 or DN150 Drop antennas 2 Radius of the silo r 3 The minimum measured level for...

Страница 25: ...They can be installed adjacent to other radar level transmitters Point the device in the correct direction to get the best performance Figure 3 5 Point the device in the correct direction to get the...

Страница 26: ...oms Dish shaped or conical bottoms have an effect on the measuring range The device cannot measure to the bottom of the tank If possible install the device as shown in the illustration that follows Fi...

Страница 27: ...ey have an effect on the performance of the device WARNING LPR Level Probing Radar LPR Level Probing Radar LPR Level Probing Radar LPR Level Probing Radar devices measure level in the open air or in a...

Страница 28: ...ctrum recording on page 100 2 Beam radius of the antenna refer to the table below The beam radius increases by increments of x mm for each metre of distance from the antenna Antenna type Beam angle Be...

Страница 29: ...ers the silo touches the antenna the device will measure incorrectly If the product fills the silo directly below the antenna the device will also measure incorrectly For more data about the measuring...

Страница 30: ...e sure that you point the device in the correct direction For more data refer to General notes on page 24 Point the device in the correct direction Tighten the flange bolts Refer to local rules and re...

Страница 31: ...len wrench Metallic Horn antennas without an antenna extension Metallic Horn antennas without an antenna extension Metallic Horn antennas without an antenna extension Metallic Horn antennas without an...

Страница 32: ...8 1 G connection ISO 228 1 G connection ISO 228 1 G connection ISO 228 1 G connection Align the gasket correctly NPT connection NPT connection NPT connection NPT connection Wind the thread seal tape a...

Страница 33: ...connection NPT connection NPT connection Wind the thread seal tape around the process connection in agreement with good engineering practice Put the device carefully on the silo process connection Do...

Страница 34: ...antenna extensions If the device is not supplied with the antenna attached to the signal converter or antenna extensions are supplied as an accessory after delivery of the device do the procedures tha...

Страница 35: ...ge the value Push to change the position of the cursor Push to decrease the value or to increase the value i Each antenna extension is 105 mm long If the device has 3 antenna extensions they have a to...

Страница 36: ...settings for a device with an antenna extension Push 2 2 and to go to menu item Login Login Login Login Enter the hexadecimal password default password 0058 Push 2 2 2 5 and to go to menu item C 1 8 A...

Страница 37: ...ion 50 mm open end wrench not supplied G 3 or 3 NPT connection G 3 or 3 NPT connection G 3 or 3 NPT connection G 3 or 3 NPT connection 65 mm open end wrench not supplied 3 mm Allen wrench not supplied...

Страница 38: ...i End of the procedure Equipment needed Device Orientation system accessory ISO 228 1 G connection ISO 228 1 G connection ISO 228 1 G connection ISO 228 1 G connection Gasket not supplied NPT connect...

Страница 39: ...n end wrench NPT connection NPT connection NPT connection NPT connection Wind the thread seal tape around the process connection of the device in agreement with good engineering practice Put the devic...

Страница 40: ...get a good measurement signal You can tilt the device to a maximum angle of 30 from the vertical axis Equipment needed 5 mm Allen wrench not supplied Loosen the 4 socket head screws around the orient...

Страница 41: ...s it difficult to read the display you can rotate the display in increments of 90 Equipment needed Cover wrench Display extractor Follow this procedure Remove the housing cover with the cover wrench F...

Страница 42: ...k Make sure that the housing cover has a gasket Attach the cover to the housing and tighten it by hand i End of the procedure 3 11 Weather protection 3 11 1 How to attach the weather protection to the...

Страница 43: ...weather protection clamp Tighten the lock ing nuts with a 10 mm socket wrench 3 Lower the weather protection cover onto weather protection clamp until the hole for the lock is in the slot at the fron...

Страница 44: ...2 How to open the weather protection 1 Remove the R clip from the hole at the front of the weather protection cover 2 Remove the weather protection cover 3 Lift the display screen cover End of the pr...

Страница 45: ...rk done on the electrical components of the measuring device may only be carried out by properly trained specialists Look at the device nameplate to ensure that the device is delivered according to yo...

Страница 46: ...cable is shielded 2 Current output 3 Current output 4 Location of the external grounding terminal at the bottom of the converter Electrical power to the output terminal energizes the device The output...

Страница 47: ...ent needed 3 mm Allen wrench not supplied Cover wrench Procedure 1 Loosen the lock screw with a 3 mm Allen wrench 2 Remove the cover stop 3 Turn the cover counterclockwise with the cover wrench 4 Remo...

Страница 48: ...edure 1 Loosen the cable gland Put the electrical wires into the cable entry Loosen the terminal screws with a POZIDRIV PZ1 screwdriver Connect the electrical wires to the connector 2 Tighten the term...

Страница 49: ...t needed 3 mm Allen wrench not supplied 1 Put the cover on the housing 2 Turn the cover clockwise until it is fully engaged 3 Attach the cover stop and lock screw 4 Tighten the lock screw with a 3 mm...

Страница 50: ...current output 4 Pin 4 not connected 5 Grounding terminal external thread of the connector 6 Location of the external grounding terminal at the bottom of the converter Electrical power to the output t...

Страница 51: ...ummy plugs 3 The diameter of the outer sheath of the electrical cable for the power supply and current output must be 6 10 mm or 0 24 0 39 DANGER For electrical data for device operation in hazardous...

Страница 52: ...twork The device output is factory set to communicate point to point To change the communication mode from point to point point to point point to point point to point to multi drop multi drop multi dr...

Страница 53: ...ti drop networks Figure 4 9 Multi drop network non Ex 1 Address of the device each device must have a different address in multidrop networks 2 4 mA HART 3 Resistor for HART communication typically 25...

Страница 54: ...ce will show measurement data Measurements agree with specifications given in the customer order 5 3 Operating concept You can read measurements and configure the device with A digital display screen...

Страница 55: ...Meas page or C6 5 1 Function Function Function Function 2nd Meas Page If menu item C6 4 2 1st Value Variable 1st Meas page is set to Level then the device shows Level as the current output percentage...

Страница 56: ...ormal mode Normal mode Normal mode None Program mode Program mode Program mode Program mode Menu Menu Menu Menu Exit the menu If you are in the top level menu the device goes back to Normal mode Menu...

Страница 57: ...ied with the device 2 Push the buttons on the keypad i This will operate the device Equipment needed Bar magnet How to push the keypad buttons with your hand Figure 5 3 How to push the keypad buttons...

Страница 58: ...for sending information between the system and the field device This standard agrees with IEC 62453 Field devices are easily integrated Installation is supported by a user friendly Wizard Install the...

Страница 59: ...ger is an industrial Plant Asset Management PAM software tool Its role is to Store configuration information for each device Store and read process data Store and read diagnostic status information He...

Страница 60: ...mode Use this mode to read and change parameters commission the device create tables for volume or mass measurement and change critical values to measure in difficult process conditions To change set...

Страница 61: ...LCD display settings When you do the Standard setup Standard setup Standard setup Standard setup procedure the measurement value on the first measurement page of the LCD display in Normal mode automat...

Страница 62: ...digits and decimal places in the measurement values shown in How to change the number of digits and decimal places in the measurement values shown in How to change the number of digits and decimal pla...

Страница 63: ...p Full Setup Full Setup Conversion Conversion Conversion Conversion For data on how to prepare the conversion table refer to How to configure the device to measure volume or mass on page 107 m L hL he...

Страница 64: ...the C Full Setup C Full Setup C Full Setup C Full Setup menu to change settings You can set up a conversion table for volume or mass measurement change current output values change HART settings chang...

Страница 65: ...able at the User access level Normal mode and menu items B2 Actual Values and C7 3 1 Message View Change All sub menus in menus A Quick Setup B Test and C Full Setup NOTE NOTE NOTE NOTE You can change...

Страница 66: ...mber 1 9 or a letter A F If you change the password for the Expert access level then the first two digits must be zero 00xx The last two digits can be a number 1 9 or a letter A F Enter the new passwo...

Страница 67: ...rogram mode Push 2 to go to the menu item A3 Login Push Enter the password used at this time for the Expert access level If it is the default password enter 0058 Push and to go to the menu item A4 1 S...

Страница 68: ...to sub menu C1 1 Enter the menu item Enter Go up a menu level for example from sub menu C1 1 to menu C1 Go to Normal mode If you changed settings in Program mode you must save or cancel your new sett...

Страница 69: ...Push to enter the menu item 2 Push or to change the parameter 3 Push to set the new parameter and go back to the menu level 4 Parameter 5 Menu item name 6 The factory default value left side and fact...

Страница 70: ...and min max symbol right side 8 A tick symbol shows that there is a new setting the new setting is not saved at this time If menu items have values that you can change very large and very small value...

Страница 71: ...n from Yes Yes Yes Yes No No No No or Back Back Back Back Back Back Back Back makes the display go back to Program mode If this screen is set to Yes or No push to accept Yes or reject No the new setti...

Страница 72: ...eight A4 1 3 1 Current Out 1 Var A4 1 3 2 0 Range A4 1 3 3 100 Range A4 1 3 4 Current Out Range A4 1 3 5 Error Function A4 2 Empty A4 2 Empty A4 2 Empty A4 2 Empty Spectrum Spectrum Spectrum Spectrum...

Страница 73: ...ues Values Values Values B2 1 Operating Time B2 3 Sensor Value B2 4 Level B2 5 Distance B2 6 Reflection B2 7 Level Lin 2 B2 8 Volume 2 B2 9 Mass 2 B2 10 Distance Lin 2 B2 11 Ullage Volume 2 B2 12 Ulla...

Страница 74: ...ld C2 7 Mult Refl Enable C2 8 Empty Spectrum Enable C3 Conversion C3 Conversion C3 Conversion C3 Conversion 1 C3 1 Edit Table C3 1 Edit Table C3 1 Edit Table C3 1 Edit Table C3 1 1 Erase Table 2 C3 1...

Страница 75: ...2 7 Identification Universal Revision C5 1 2 8 Identification Device Revision C5 1 2 9 Identification Software Revision C5 1 2 10 Identification Hardware Revision C5 1 3 1 Device Information Descript...

Страница 76: ...ormat 1st Value C6 5 6 2nd Value Variable 6 C6 5 7 Format 2nd Value 6 C6 5 8 3rd Value Variable 6 C6 5 9 Format 3rd Value 6 C7 Device C7 Device C7 Device C7 Device C7 1 Information C7 1 Information C7...

Страница 77: ...C Full Setup menu C6 5 2nd Meas Page in the section that follows 7 This menu item can only be unlocked at the factory Menu No Function Function description Selection list Default A1 Language Make a se...

Страница 78: ...through this filter to correctly measure the tank contents This menu item starts a quick set up procedure We recommend that the tank is empty or only filled to the minimum level before you do the proc...

Страница 79: ...istance value This procedure sends an output signal that agrees with the test reading Output will change to the selected value independent of the measurement data When you push to confirm the value th...

Страница 80: ...a volume value This procedure sends an output signal that agrees with the test reading Output will change to the selected value independent of the measurement data When you push to confirm the value t...

Страница 81: ...device simulates an ullage volume value This procedure sends an output signal that agrees with the test reading Output will change to the selected value independent of the measurement data When you p...

Страница 82: ...measured is less than 1 metre then the reflection value is 0 Read only B2 7 Level Lin This menu item shows the level readings linearized measured at this time This measurement data is shown with the u...

Страница 83: ...er Spectrum This menu item shows the raw spectra data of the radar signal which makes a reflection on the surface of the silo contents and is received by the device This data is shown as a graph of th...

Страница 84: ...show the blocking distance on the display screen Minimum access level to change the Minimum access level to change the Minimum access level to change the Minimum access level to change the setting se...

Страница 85: ...to change the Minimum access level to change the Minimum access level to change the setting setting setting setting Expert min max 5000 0 0 5000 0 m 196 85 03 196 85 03 16404 16404 ft 0 m 0 ft C2 Proc...

Страница 86: ...evel to change the setting setting setting setting Expert Disabled Enabled Disabled 1 C2 6 Overfill Threshold If you have set menu item C2 5 Overfill Detection to Enabled this menu item will be availa...

Страница 87: ...imum access level to change the Minimum access level to change the Minimum access level to change the setting setting setting setting Expert Volume Mass Linearization Volume C3 2 Input Table C3 2 Inpu...

Страница 88: ...1 06 mm 192 91 03 200 79 03 16076 16732 ft C1 2 Tank Height C1 5 Blocking Distance 1 C4 1 4 Current Out Range This menu item sets the limits of the output current range to 1 of 4 available options sta...

Страница 89: ...Enter the measured value Minimum access level to change the Minimum access level to change the Minimum access level to change the Minimum access level to change the setting setting setting setting Exp...

Страница 90: ...ge the Minimum access level to change the Minimum access level to change the Minimum access level to change the setting setting setting setting Operator C5 1 3 2 Message You can give more data in this...

Страница 91: ...trast control for the display screen You can select a shade of grey between light grey 10 and black 10 Minimum access level to change the Minimum access level to change the Minimum access level to cha...

Страница 92: ...setting User Level Distance Sensor Value Reflection 2 Level C6 4 7 Format 2nd Value You can change the number of decimal places that are shown for the second value on the measurement page in Normal mo...

Страница 93: ...mum access level to change the Minimum access level to change the Minimum access level to change the Minimum access level to change the setting setting setting setting User Refer to Function descripti...

Страница 94: ...ly name and model code Read only C7 1 4 V Number This is the manufacturer s option code for device configuration Read only C7 1 5 Electronic Revision This is the hardware revision number This number a...

Страница 95: ...u set this menu item to Cst custom length unit enter values in menu items C7 5 2 1 thru C7 5 2 3 Minimum access level to change the Minimum access level to change the Minimum access level to change th...

Страница 96: ...ange kg kilogram to the custom mass unit 1 0 C7 6 Factory Default C7 6 Factory Default C7 6 Factory Default C7 6 Factory Default C7 6 1 Reset to Fact Def If you set this menu item to YES the device go...

Страница 97: ...ters and values as the current output settings measurement variable 0 range and 100 range The first measurement page shows one value and bar by default If you must show one or more measurement values...

Страница 98: ...is necessary to measure volume ullage volume mass or ullage mass refer to How to configure the device to measure volume or mass on page 107 to change the position of the cursor to decrease the value...

Страница 99: ...he measuring range between 0 and 100 values given in the 0 range and 100 range steps If you set this menu item to 4 20 mA or 3 8 20 5 mA 0 Range 4 mA 100 Range 20 mA If you set this menu item to 4 20...

Страница 100: ...s on when menu item C2 8 Empty Spectrum Enable is set to Enable it will ignore the interference signals Before you do the empty spectrum recording procedure install the device on the tank For more dat...

Страница 101: ...pe Type Type to Partial Average or Partial Max you will have one more step to do in this procedure You must give a partial distance value less than or equal to the distance to the product surface from...

Страница 102: ...2 to go to menu item C5 1 2 1 Polling Address Enter the password used at this time for the Expert access level If it is the default password enter 0058 Push to confirm Enter the value 000 and push to...

Страница 103: ...ge facing is the reference point for distance measurement 0 m 0 ft 0 The position of the measurement scale specified by the 0 Range and 100 Range settings is related to this reference point If you con...

Страница 104: ...mode Bar graph function in Normal mode There is an optional bar graph shown on the two measurement pages in Normal mode set C6 4 1 C6 5 1 Function to One Value Bar or Two Values Bar By default the bar...

Страница 105: ...ft 0 The position of the measurement scale specified by the 0 Range and 100 Range settings is related to this reference point If you configured a measurement scale for the current output signal in the...

Страница 106: ...unction in Normal mode Bar graph function in Normal mode There is an optional bar graph shown on the two measurement pages in Normal mode set C6 4 1 C6 5 1 Function to One Value Bar or Two Values Bar...

Страница 107: ...1 Edit Table to delete strapping table data that the device uses at this time Erase Table Push and to set this menu item to Yes to delete the data Push and then and to make a selection from the conver...

Страница 108: ...he device will ask you to save or cancel your settings Push or to make a selection from Yes Yes Yes Yes No No No No or Back Back Back Back Back Back Back Back makes the display go back to Program mode...

Страница 109: ...be a minimum of 20 more than the true tank height Go to a measurement page in Normal mode that shows the distance Make a note of the distance of the reflection measured by the device i This is the de...

Страница 110: ...he next step We recommend that you do an empty spectrum scan when the silo is empty and all the moving parts are in operation Figure 6 10 How to make a filter to remove radar signal interference 1 Emp...

Страница 111: ...ymbol is shown with the message Status messages NAMUR NE 107 Normal mode device status symbol If the device status changes the device shows a status symbol in the top left corner of the display screen...

Страница 112: ...button to show more data If the device shows an error message refer to the table that follows for more data and to find a solution to a problem Figure 6 12 Device status Normal mode device status pag...

Страница 113: ...device again If the message is shown again tell the supplier Fatal Converter Error CO 2 An electronics or hardware failure occurred Fatal Converter Error Generic An electronics or hardware failure oc...

Страница 114: ...Revision Revision Revision If it is necessary tell the supplier Conv NVRAM Layout Error Incorrect data in the parameter memory De energize and then energize the device again If the message is shown ag...

Страница 115: ...nsor Sensor Voltage Low The voltage supplied to the sensor is too low Do a check of the power supply Strong Signal The signal amplitude is too large It is possible that the signal is saturated Make su...

Страница 116: ...check of the process conditions and the maximum current output value M M M M Sensor Sensor Sensor Sensor Weak Signal The signal amplitude is too small Make sure that the device installation agrees wi...

Страница 117: ...the device Peak Lost in Tank Bottom The signal was lost near the tank bottom Spherical or conical tanks can cause the device to show this status message The device will measure the level of tank cont...

Страница 118: ...he device obey the instructions given in the handbook 7 1 2 Maintenance of the O rings for the housing covers If the display cover 1 or the terminal compartment cover 2 of the housing is opened and cl...

Страница 119: ...more data refer to the table that follows CAUTION Use multi purpose grease that is applicable for the operating temperature range of the O ring with the properties that follow Operating temperature r...

Страница 120: ...subject to wear and tear under normal operating conditions Process conditions How to use the purge The device is colder than other elements in the process It is possible that there will be condensati...

Страница 121: ...strict attention to the following points Due to statutory regulations on environmental protection and safeguarding the health and safety of the personnel the manufacturer may only handle test and repa...

Страница 122: ...Date Signature Stamp LEGAL NOTICE Disposal must be carried out in accordance with legislation applicable in your country Separate collection of WEEE Waste Electrical and Electronic Equipment in the Eu...

Страница 123: ...e product For further signal processing the difference f is calculated from the actual transmitted frequency and the received frequency The difference is directly proportional to the distance A large...

Страница 124: ...lect Direct Plus mode If you select Direct Plus mode the device locks on the level signal and monitors changes in level If the device then finds larger reflections in the silo it will only monitor the...

Страница 125: ...ment system consists of a measuring sensor antenna and a signal converter Options Integrated LCD display 20 70 C 4 158 F if the ambient temperature is not in these limits then this condition can stop...

Страница 126: ...air humidity 60 15 Target Metal plate in an anechoic chamber The device has specified settings Operating conditions Temperature Temperature Temperature Temperature Ambient temperature 40 80 C 40 176 F...

Страница 127: ...allation on page 21 Dimensions and weights For dimensions and weights data refer to Dimensions and weights on page 136 Materials Housing Polyester coated aluminium Option Stainless steel 1 4404 316L n...

Страница 128: ...Maximum current 21 5 mA Current output load Non Ex Ex i Non Ex Ex i Non Ex Ex i Non Ex Ex i RL Uext 12 V 21 5 mA For more data refer to Minimum power supply voltage on page 133 Ex d Ex d Ex d Ex d RL...

Страница 129: ...nically isolated Communication standard H1 ITK version 6 3 Function blocks 1 Enhanced Resource Block RB 1 Customer Level Transducer Block LEVELTB 1 Customer Converter Transducer Block CONVTB 1 Custome...

Страница 130: ...vision ratings Division ratings Division ratings XP IS Class I Div 1 GPS ABCD T6 Tx DIP Class II III Div 1 GPS EFG T85 C T C 4 IS Class I Div 1 GPS ABCD T6 Tx IS Class II III Div 1 GPS EFG T85 C T C 4...

Страница 131: ...vel for the Failure Information of Digital Transmitters NAMUR NE 53 Software and Hardware of Field Devices and Signal Processing Devices with Digital Electronics NAMUR NE 107 Self Monitoring and Diagn...

Страница 132: ...mum recommended blocking distance antenna extension length antenna length 300 mm Figure 8 3 Measuring accuracy graph of measuring accuracy in inches against measuring distance in ft X Measuring distan...

Страница 133: ...ved Ex i IS devices Figure 8 4 Minimum power supply voltage for an output of 21 5 mA at the terminals Non Ex and Hazardous Location approval Ex i IS X Power supply U V DC Y Current output load RL Haza...

Страница 134: ...de rating EN 1092 1 flange and threaded connection in C and barg Figure 8 7 Pressure temperature de rating EN 1092 1 flange and threaded connections in F and psig 1 Process pressure p barg 2 Process c...

Страница 135: ...vices that are installed on a pressure vessel and used in Canada Figure 8 8 Pressure temperature de rating ASME B16 5 flange and threaded connections in C and barg Figure 8 9 Pressure temperature de r...

Страница 136: ...readed connections Figure 8 10 Metallic Horn antennas with G or NPT threaded connections The diameter of the outer sheath of the cable must be 7 12 mm or 0 28 0 47 Cable glands for cQPSus approved dev...

Страница 137: ...200 8 151 185 662 1 847 1 190 1 This is the dimension without the antenna extension option A maximum of 10 antenna extensions are available Each antenna exten sion is 105 mm long Horn antenna version...

Страница 138: ...ange connections 1 Metallic Horn antenna with a flange connection 2 Metallic Horn antenna with a flange connection and the 2 slanted flange option The diameter of the outer sheath of the cable must be...

Страница 139: ...es without the antenna extension option A maximum of 10 antenna extensions are avail able Each antenna extension is 105 mm long 2 If the device has the 2 slanted flange option Horn antenna version Dim...

Страница 140: ...n antenna with a low pressure flange attached to a G threaded connection ISO 228 1 2 Metallic Horn antenna with a low pressure flange attached to an NPT threaded connection ASME B1 20 1 The diameter o...

Страница 141: ...8 151 215 247 657 1 847 1 904 1 190 1 This is the dimension without the antenna extension option A maximum of 10 antenna extensions are available Each antenna exten sion is 105 mm long Horn antenna v...

Страница 142: ...nnections Figure 8 13 Drop antennas with threaded connections The diameter of the outer sheath of the cable must be 7 12 mm or 0 28 0 47 Cable glands for cQPSus approved devices must be supplied by th...

Страница 143: ...5 221 1 406 1 144 1 This is the dimension without the antenna extension option A maximum of 5 antenna extensions are available Each antenna extension is 105 mm long Drop antenna version Dimensions inc...

Страница 144: ...nge connection 2 Drop antenna with a flange connection and a flange plate protection option 3 Drop antenna with a flange connection and a 2 slanted flange option The diameter of the outer sheath of th...

Страница 145: ...e PP or PTFE flange protection option then f 20 mm If the device has the 2 slanted flange option then f 10 mm 3 If the device has the 2 slanted flange option then g 2 Drop antenna version Dimensions i...

Страница 146: ...h a low pressure flange attached to a G threaded connection ISO 228 1 2 Metallic Horn antenna with a low pressure flange attached to an NPT threaded connection ASME B1 20 1 The diameter of the outer s...

Страница 147: ...nsions mm a b c d e G NPT G NPT DN80 3 5 94 7 40 8 66 5 35 1 12 76 1 14 01 1 2 91 DN100 4 5 94 7 40 8 66 6 26 1 13 66 1 14 92 1 3 70 DN150 6 5 94 7 40 8 66 8 54 1 15 94 1 17 20 1 5 67 1 This is the di...

Страница 148: ...n option Figure 8 17 Weather protection option 1 Front view with weather protection closed 2 Left side with weather protection closed 3 Rear view with weather protection closed Dimensions Weights kg a...

Страница 149: ...6 8 130 9 DN100 PTFE Drop antenna with process connection standard length 1 3 8 60 2 8 4 132 7 DN150 PTFE Drop antenna with process connection standard length 1 7 2 63 6 15 8 139 9 Antenna extension o...

Страница 150: ...sions for use in hazardous areas There are 2 primary operation modes for HART compatible devices point to point mode and multi drop mode If the device is used in point to point mode the HART Protocol...

Страница 151: ...ed There are two ways of using the HART communication as Point to Point connection and as Multi Drop connection with 2 wire connection 9 3 1 Point to Point connection analogue digital mode Point to Po...

Страница 152: ...undation Fieldbus devices Device Descriptions DDs are used to integrate different devices into the Field Communicator 9 5 1 Installation System and software requirements for the Field Communicator Sys...

Страница 153: ...nstallation Kit HART AMS You can download this exe file from our website For installation data refer to the AMS Intelligent Device Manager Books Online section Basic AMS Functionality Device Configura...

Страница 154: ...ation data refer to the documentation that is supplied with the Device DTM in the Downloads section of the website 9 7 2 Operation The DTM and the device s local display use almost the same procedures...

Страница 155: ...Product Epsilon R Gas Measuring Mode Overfill Detection Overfill ThresholdOpt Mult Refl Enable Empty Spectrum Enable Min Peak Required Min Plausibility Window Conversion Select Conversion Entry Count...

Страница 156: ...it Volume Unit Mass Log Operating TimeRd Reset Log Messages Factory Default Reset To Fact Default Full Setup HART Identification Info Loop current Mode Online Mode Poll Address Tag Long Tag Manufactur...

Страница 157: ...evel LinearizationRd Opt VolumeRd Opt MassRd Opt Distance LinearizationRd Opt UllageRd Opt Ullage MassRd Opt Sensor TempRd Converter TempRd Simulation Process Variables Simulation Values Device Status...

Страница 158: ...4 ATEX II 3 G Ex nA T6 T4 Gc 5 NEPSI Ex ia IIC T6 T4 Ga Gb Ex iaD 20 21 T85 C T100 C or T85 C T130 C IP6X 6 NEPSI Ex d ia IIC T6 T4 Ga Gb Ex iaD tD A20 A21 T85 C T100 C or T85 C T130 C IP6X A cQPSus...

Страница 159: ...cover without window 4 Plug in display cover with window Display Display Display Display Documentation language Documentation language Documentation language Documentation language 1 English 2 German...

Страница 160: ...17 316L 5 525 mm 21 316L 6 630 mm 24 316L for Metallic Horn antennas 7 735 mm 29 316L for Metallic Horn antennas 8 840 mm 33 316L for Metallic Horn antennas A 945 mm 37 316L for Metallic Horn antennas...

Страница 161: ...0 PN16 Type B1 ASME B16 5 flange L 1 A 3 150 lb RF L 2 A 3 300 lb RF M 1 A 4 150 lb RF M 2 A 4 300 lb RF P 1 A 6 150 lb RF P 2 A 6 300 lb RF R 1 A 8 150 lb RF JIS B2220 flange L U P 80A JIS 10K RF M U...

Страница 162: ...ies Tag plate Accessories Tag plate Accessories Tag plate Accessories Tag plate 0 Without 1 Weather protection 3 Stainless steel Tag plate 18 characters max 6 Weather protection Stainless steel Tag pl...

Страница 163: ...FDX070500 Metallic Horn DN200 8 1 XFDX070600 PP Drop DN80 3 1 XFDX070700 PP Drop DN100 4 1 XFDX070800 PP Drop DN150 6 1 XFDX070900 PTFE Drop DN80 3 1 XFDX071000 PTFE Drop DN100 4 1 XFDX071100 PTFE Dro...

Страница 164: ...r display cover 9 or blind cover 11 9 Display cover 10 Dummy module 11 Blind cover 12 Hinge 13 Plastic sun cover for display cover 9 Description Quantity Part reference Housing Aluminium blind cover 1...

Страница 165: ...Non Ex GP or Ex i IS 4 4 4 4 1 XFDX040100 Terminal cartridge Ex d XP 4 4 4 4 1 XFDX040200 Cable gland Cable entry Cable gland M20 1 5 Plastic black non Ex GP 5 5 5 5 10 XFDX050100 Cable gland M20 1 5...

Страница 166: ...10 2 Accessories general items 1 Weather protection 2 Viator RS232 HART converter 3 Viator USB HART converter 4 Display extractor magnet and cover wrench Item Description Quantity Part reference 1 We...

Страница 167: ...tion on level transmitter 1 Disc with holes drilled according to DN50 PN2 5 PN40 2 150 lb 1 XF70000015 Disc with bolt holes drilled according to DN80 PN2 5 PN40 3 150 lb XF70000016 Disc with holes dri...

Страница 168: ...C 392 F 1 XF70000028 DN100 PN16 PP max 100 C 212 F 1 XF70000024 DN100 PN16 PTFE max 150 C 392 F 1 XF70000029 DN100 PN40 PP max 100 C 212 F XF70000025 DN150 PN16 PP max 100 C 212 F 1 XF70000026 DN150 P...

Страница 169: ...oles drilled according to DN150 PN2 5 PN40 6 150 lb XF70000N63 Disc with holes drilled according to DN200 PN2 5 PN40 8 150 lb XF70000N64 Other accessories for process connections 5 316L wall mounted b...

Страница 170: ...Manager A driver for use in the PACTware program All data and functions of the device are included in it Electromagnetic compatibility Electromagnetic compatibility Electromagnetic compatibility Elec...

Страница 171: ...lection Radar reflection Radar reflection Radar reflection Signal reflected from the surface of the silo contents Signal converter Signal converter Signal converter Signal converter A set of electroni...

Страница 172: ...adar Manual v1 00 en Figure 10 4 Measurement definitions distance 1 Distance 2 Blocking distance 3 Flange facing 4 Gas Air 5 Tank height 6 Ullage volume or mass Figure 10 5 Measurement definitions lev...

Страница 173: ...NOTES 11 173 Senator S24S 01 2019 Senator S24S Radar Manual v1 00 en www hawkmeasurement com...

Страница 174: ...11 NOTES 174 Senator S24S 01 2019 Senator S24S Radar Manual v1 00 en www hawkmeasurement com...

Страница 175: ...NOTES 11 175 Senator S24S 01 2019 Senator S24S Radar Manual v1 00 en www hawkmeasurement com...

Страница 176: ...5 17 Maurice Court Nunawading VIC 3131 AUSTRALIA Phone 61 3 9873 4750 Fax 61 3 9873 4538 info hawk com au Hawk Measurement 5010 Gateway Drive Medina OH 44256 USA Phone 1 888 429 5538 Phone 1 978 304 3...

Отзывы: