6
OPERATION
80
Senator S24S
01/2019 - Senator S24S Radar Manual v1.00 en
B1.1.5
Level Lin.
This sets the device to a given test level
value (linearized). This menu item is
only available if you set up a
linearization table in menu item C3.1
C3.1
C3.1
C3.1
Edit Table
Edit Table
Edit Table
Edit Table. Make sure that menu item
C4.1.1 Current Out. 1 Var. is set to "Lin.
Level" before the device simulates a
level value. This procedure sends an
output signal that agrees with the test
reading. Output will change to the
selected value, independent of the
measurement data.
When you push [
^
^
^
^
] to confirm the
value, the display shows the question
"Start Simulation?" Push [
] or [
] to
set the display to "Yes". Push [
^
^
^
^
] again
to start the test. The device goes to
back to Normal mode after 1 hour.
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
procedure:
procedure:
procedure:
procedure: Expert
min-max:
-5000.0...+5000.0 m /
-196.85
+03
...+196.85
+03
in /
-16404...+16404 ft
1
B1.1.6
Volume
This sets the device to a given test
volume value. This menu item is only
available if you set up a conversion
(volume) table in menu item C3.1 Edit
C3.1 Edit
C3.1 Edit
C3.1 Edit
Table
Table
Table
Table. Make sure that menu item
C4.1.1 Current Out. 1 Var. is set to
"Volume" before the device simulates a
volume value. This procedure sends an
output signal that agrees with the test
reading. Output will change to the
selected value, independent of the
measurement data.
When you push [
^
^
^
^
] to confirm the
value, the display shows the question
"Start Simulation?" Push [
] or [
] to
set the display to "Yes". Push [
^
^
^
^
] again
to start the test. The device goes to
back to Normal mode after 1 hour.
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
procedure:
procedure:
procedure:
procedure: Expert
min-max:
0...1.00
+06
m
³
1
B1.1.7
Mass
This sets the device to a given test
mass value. This menu item is only
available if you set up a conversion
(mass) table in menu item C3.1 Edit
C3.1 Edit
C3.1 Edit
C3.1 Edit
Table
Table
Table
Table. Make sure that menu item
C4.1.1 Current Out. 1 Var. is set to
"Mass" before the device simulates a
mass value. This procedure sends an
output signal that agrees with the test
reading. Output will change to the
selected value, independent of the
measurement data.
When you push [
^
^
^
^
] to confirm the
value, the display shows the question
"Start Simulation?" Push [
] or [
] to
set the display to "Yes". Push [
^
^
^
^
] again
to start the test. The device goes to
back to Normal mode after 1 hour.
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
Minimum access level to do the
procedure:
procedure:
procedure:
procedure: Expert
min-max:
0...10.000
+09
kg
1
Menu
No.
Function
Function description
Selection list