82
Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
26.Press
YES
to continue. The screen displays:
27.Press
YES
to continue. The screen displays:
28.Press
YES
to continue. The screen displays:
Check, Press YES To Continue
Bowl is Seated Correctly in the
Centrifuge.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
Additive Solution Seal
is broken.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
Sterile connect RBC bag to RED
Striped tubing.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240