114
Calibration and Configuration
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
19.Press
MODIFY PROGRAM
to move the pointer on the screen to the De-
Glycerolization protocol option parameter.
20.Press START. The screen displays:
Figure 34, Bowl size option in deglycerolization
21.Press
MODIFY PROGRAM
, if necessary, to move the pointer on the
screen to the
Bowl size option
parameter.
22.Press START. The screen displays:
23.Press
MODIFY PROGRAM
to change the of the Bowl Size setting within
the deglycerolization protocol if necessary, then press
START
to
confirm.
24.Press STOP five times to return to the
Main Configuration
screen.
GLYCEROLIZATION CONFIGURATION
International
Cell Wash
Glycerolization
-> De-Glycerolization
DEGLYCEROLIZATION CONFIGURATION
International
-> Bowl size option
BOWL SIZE OPTION PARAMETER
-> Small Bowl Only (275ml)
Large Bowl Only (325ml)
Both
Status = Small Bowl Only
CONFIGURATION
-> PROTOCOLS CONFIGURATION
LANGUAGE
MACHINE
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......