Preventive Maintenance and Installation
97
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
The following display appears:
3. Wait until compressor stops.
4. Verify that the compressor does not turn on during 5 minutes.
5. If this is not the case check the pneumatic circuit to find and repair the
leakage then test again.
DPM and
SPM Sensor
Test
Note: Power on the device in the Utilities mode of operation by pressing and
holding
SHAKER
while switching the power on. Hold
SHAKER
until the
UTILITIES menu appears on the display. On the UTILITIES menu, press
MODIFY PROGRAM
until the arrow points to DIAGNOSTICS, then press
START
.
1. Be sure that no filters are connected to the DPM and SPM. Select the
pressure test using
MODIFY PROGRAM
and
START
.
2. Verify that the DPM/SPM offsets are within the range of 1902–2194. If
out of range, DPM/SPM may need to be replaced.
3. Grasp the DPM white luer plunger and twist counterclockwise while
sliding the plunger in. Verify both interlock switch status indicators
change to IN. Release the plunger and verify that both indicators return
to OUT.
4. Grasp the DPM white luer plunger and twist clockwise while sliding the
plunger in. Verify both interlock switch status indicators change to IN.
Release the plunger and verify that both indicators return to OUT.
5. Grasp the SPM white luer plunger and twist counterclockwise while
sliding the plunger in. Verify both interlock switch status indicators
change to IN. Release the plunger and verify that both indicators return
to OUT.
6. Grasp the SPM white luer plunger and twist clockwise while sliding the
plunger in. Verify both interlock switch status indicators change to IN.
Release the plunger and verify that both indicators return to OUT.
7. Install a new DPM/SPM filter with about 4” of tubing onto the DPM.
Connect a calibrated electronic/digital measuring device or equivalent
device to the filter tubing. Pressurize the DPM to 1/- 5mmHg
and verify that the displayed pressure reading is within the range of
95–105 mmHg.
VALVES
VALVE 1
O
E
L
VALVE 4
O
E
L
VALVE 2
O
E
L
VALVE 5
O
E
L
VALVE 3
O
E
L
VALVE 6
O
E
L
PNEU P :
XXXX
INTERVAL:
8
ERRORS:
0
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......