Calibration and Configuration
121
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
6. Press
MODIFY PROGRAM
until the arrow points to RS232 Device and
then press
START
. The following screen is displayed:
7. Press
MODIFY PROGRAM
until printer is selected, then press
START
.
The following screen is displayed:
8. Press
STOP
three times to exit. The following screen appears:
9. Switch the memory protect switch on the CPU board to the Protect On
mode (up) and the following display appears:
10.Press
STOP
and the device restarts and performs the self test.
Device
Data Logging
Note: Data logging, recording of procedures, is not an active option on the
device at this time.
RS232 Device
—> Disk Drive
Printer
Device = XXXXX
RS232 Device
Disk Drive
—> Printer
Device = XXXXX
CONFIGURATION
PROTOCOLS CONFIGURATION
—> LANGUAGE
******************************
**
REPLACE MEMORY
**
**
PROTECTION
**
******************************
UTILITIES
CALIBRATION
DIAGNOSTICS
—> CONFIGURATION
MAINTENANCE