Preventive Maintenance and Installation
75
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Figure 19, Printer controls
3. On the UTILITIES menu, press
MODIFY PROGRAM
until the arrow
points to MAINTENANCE, then press
START
.
4. Press
MODIFY PROGRAM
until the arrow points to PRINT CALIB, then
press
START
.
5. Press
MODIFY PROGRAM
until the arrow points to PRINT CALIB
VALUES, then press
Start
.
6. Verify that the printer prints out the calibration values.
Functional/
Pump/Shaker
Check
Functional Test
1. Power on the device.
After completion of system tests and cover lock test, the screen displays:
HAEMONETICS ACP 215
PLEASE SELECT PROTOCOL
-> GLYCEROLIZATION
DE-GLYCEROLIZATION